Overview of theTest Set’s Built-In Tools

Using the RF Tools Program

4.Position the cursor at Max expected loss in dBand enter the greatest loss expected through your device. The insertion loss will be displayed from 0 dB to the value entered in this field. Therefore, if the Max expected loss in dBis set higher than the measured insertion loss, the plot will not display any test data. Enter the appropriate value.

5.Position the cursor at DUPLEX OUT level in dBm and enter the appropriate value. Unless you are measuring loss through a power sensitive device, the default level should work correctly.

6.Press k1 (Begin Tst) to begin the test.

Figure 4-11 Swept Insertion Loss Setups

Calibration Setup

Measurement Setup

 

AUDIO OUT

AUDIO IN

BASEBAND OUT

DATA

P A R A L L E L P O R T 1 5

S E R I A L 9

 

 

HI

 

LO

 

 

 

 

 

 

 

 

 

 

 

P A R A L L E L P O R T 1 6

S E R I A L 1 0

 

 

 

 

I

Q

IN

 

 

ANALOG

SCOPE

EXT SCOPE

VIDEO

 

 

 

 

S E R I A L 1 1

 

 

 

 

 

MODULATION

MONITOR

TRIG IN

OUT

 

 

 

 

 

IN

OUT

 

 

 

 

 

 

 

8935

ANT IN

DUPLEX OUT

16X

 

FRAME

EVEN

TRIGGER

 

 

 

CHIP CLOCK CHIP CLOCK

 

 

RF IN/OUT

 

 

19.6608

1.2288

CLOCK

SECOND

QUALIFIER

10 MHz

EXT

 

 

MHz OUT

MHz OUT

OUT

SYNC IN

IN

REF OUT

REF IN

 

 

AUDIO OUT

AUDIO IN

BASEBAND OUT

DATA

P A R A L L E L P O R T 1 5

S E R I A L 9

 

 

HI

 

LO

 

 

 

 

 

 

 

 

 

 

 

P A R A L L E L P O R T 1 6

S E R I A L 1 0

 

 

 

 

I

Q

IN

 

 

ANALOG

SCOPE

EXT SCOPE

VIDEO

 

 

 

 

S E R I A L 1 1

 

 

 

 

 

MODULATION

MONITOR

TRIG IN

OUT

 

 

 

 

 

IN

OUT

 

 

 

 

 

 

 

8935

ANT IN

DUPLEX OUT

16X

 

FRAME

EVEN

TRIGGER

 

 

 

CHIP CLOCK CHIP CLOCK

 

 

RF IN/OUT

 

 

19.6608

1.2288

CLOCK

SECOND

QUALIFIER

10 MHz

EXT

 

 

MHz OUT

MHz OUT

OUT

SYNC IN

IN

REF OUT

REF IN

 

ANT IN

6dB Pad

Cal Cable

sweptg.eps

DUPLEX OUT

6dB Pad

ANT IN

6dB Pad

Cal Cable

DUPLEX OUT

6dB Pad

Test Cable

or Device

Barrel

Connector

7.A calibration setup screen appears with a diagram of the necessary connections (Figure 4-11). This test requires two 6-dB pads. The pads are put on the ANT IN and DUPLEX OUT ports on the Test Set to reduce impedance mismatch error at these ports. A reference level is obtained by connecting a short calibration cable between the pads on the DUPLEX OUT an ANT IN ports. Setup the hardware, then press k1 (Proceed) or the knob to continue.

8.After the Test Set has finished calibrating, a second diagram similar to the Measurement Setup of Figure 4-11appears on the screen. Connect the test cable or device to be tested between the calibration cable and one of the pads. Press k1 (Proceed) or the knob to continue.

90

Chapter 4

C:\Spk\Ref\RefGuideRevE\Tools.fm

Page 88
Image 88