Configuration - Screens and Control Fields

Control Fields on the Configuration Screens

RF Display

This field selects the format for entering the RF generator’s and RF analyzer’s frequencies:

When Freq is selected, you enter the RF generator’s and the RF analyzer’s frequencies directly using the keypad or knob.

When Chan is selected, the RF Gen Freq and Tune Freq fields on all screens are replaced by the RF Channel field, and only the channel number is entered and displayed.

Channel tuning eliminates the need to enter transmit and receive frequencies directly into the Test Set. Once your radio’s RF channel standard is selected, you only have to enter the channel number to automatically set the RF generator and RF analyzer to the correct frequencies.

Operating Considerations

See the “RF Chan Std” on page 244 for more information.

Screen(s) Where Field is Present

INSTRUMENT CONFIGURE

RF Gen Volts

This field specifies whether you want RF voltages expressed as the voltage across a 50 ohm load, or the open circuit voltage (emf).

Operating Considerations

This setting affects the RF generator’s and the tracking generator’s amplitudes.

Screen(s) Where Field is Present

INSTRUMENT CONFIGURE