Analog Measurements - Screens and Control Fields

Analog Measurements

TX Power Measurement

Transmitter power measures RF power at the RF IN/OUT port.

 

Operating Considerations

 

Only the RF IN/OUT port can be used for measuring TX power. When

 

the Input Port is set to Ant, four dashes (- - - -) appear in place of digits

 

for this measurement.

 

Use the spectrum analyzer to measure low-level RF power (≤ 60 mW) at

 

the ANT IN port.

 

See “Input Port” on page 201 or “TX Pwr Zero” on page 221 for more

 

information.

 

Connecting a signal of >60 mW to the ANT IN port can cause

CAUTION

 

instrument damage (although internal protection circuits can typically

 

withstand a short-duration signal of several Watts). If the overpower

 

circuit is triggered (signified by a warning message at the top of the

 

screen), remove the signal from the ANT IN port, and reset using Meas

 

Reset (press and release the Shift key, then the Hold key) or turn the

 

Test Set off and on to reset it.

 

Screens on which this field is displayed

 

AF ANALYZER

 

RF ANALYZER

 

RF GENERATOR

 

FM Deviation Measurement

 

This is the default setting for this measurement field. The FM deviation

 

measurement is the frequency deviation of the carrier. This

 

measurement field is only displayed when the AF Anl Infield (on the

 

AF ANALYZER screen) is set to FM Demod or FM Mod.

 

See “Displaying AF Measurements” on page 57 for more information.

 

See also “AF Anl In” on page 185.

 

Screens on which this field is displayed

 

AF ANALYZER

 

RF ANALYZER

 

RF GENERATOR

Chapter 6

183

Page 179
Image 179
Agilent Technologies 8935 series e6380a manual TX Power Measurement