Configuration - Screens and Control Fields

Control Fields on the Configuration Screens

(Gen)-(Anl)

This field is used with the RF Offset field to specify the amount of frequency offset between the RF generator and RF analyzer.

This field is always displayed when the RF Display field is set to Freq.

See “Setting an RF Generator/Analyzer Offset” on page 69 for more information.

Screen(s) Where Field is Present

INSTRUMENT CONFIGURE

(Gen)-(Anl) (User Defined)

This field defines the receiver-transmitter frequency offset when using user-defined channel operation.

Use a positive value (such as 45 MHz) when the radio’s receive frequency is higher than the transmit frequency.

Use a negative value (such as − 45 MHz) when the radio’s receive frequency is lower than the transmit frequency.

Operating Considerations

The value of this field is only used if the RF Display field is set to Chan, and the RF Chan Std field is set to USER-DEF.

For more information see the following related topics:

“Chan Space (User Defined)” on page 229 “RF Chan Std” on page 244

“RF Display” on page 245

Screen(s) Where Field is Present

INSTRUMENT CONFIGURE

Chapter 7

235