5.If the M450H has been configured with the auto-zero valve option, connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas Inlet.’ Do not exceed 30 psig in the zero air delivery line.
NOTE
The zero gas used must be clean and free of particles. If the gas contains particulate contamination, install a particulate filter in the zero air line. You can purchase an appropriate filter from Teledyne API.
6.With the ozone delivery line pressurized, adjust the ozone flow rate using the needle valve mounted inside the front panel of of the monitor. To access the needle valve, pop out the two black plastic fasteners at the top of the front panel. The panel will now hinge down, allowing access to the needle valve. The
7.Turn on the M450H by switching the power switch on the front panel. The front panel display should light with a sequence of messages,
8.Allow about 10 minutes for the UV lamp temperature to come up to it’s
P/N 02826B1 Teledyne API Model 450H O3 Monitor Instruction Manual - Page 13