Topcon COMPUTERIZED TONOMETER, CT-80 Alignment is adjusted correctly. Sometimes the shape

Page 41
MEASUREMENT SWITCH

6 Holding the alignment dot within the inner alignment mark, slightly push the machine body toward the patient.

When the machine body approaches the target eye, the alignment bar and the “FORWARD” display appear on the monitor screen.

Alignment bar

At this moment, be careful not to catch eyelashes and eyelids within the outer align- ment mark so as to ensure correct measurements.

See descriptions about the alignment bar, “FORWARD” and “TOO CLOSE” on page 36.

7 Move the machine body back and forth, with the alignment bar as a reference, while holding the alignment dot within the inner alignment mark.

When the alignment is adjusted, the shape of the inner alignment mark changes to a

.

Alignment is not adjusted

Alignment is adjusted

8 After the alignment is adjusted, press

MEASUREMENT SWITCH .

Air is ejected for measurement, and the measurement value is displayed.

In Manual mode, measurement is done by pressingeven if the alignment is not adjusted correctly. To ensure high-precision measurements, make sure the alignment is adjusted correctly.

 

If the shape of the inner alignment mark does not change to a

MEMO

even after correctly adjusting the alignment, check again to see if

the alignment is adjusted correctly. Sometimes the shape of the

 

inner alignment mark does not change to a

if the condition of

 

the cornea is unfavorable.

 

40

BASIC OPERATIONS

Image 41
Contents CT-80 Introduction This instrument features the followingDisplay Meaning Display for Safe USEIcons Prevention item Safety CautionsIcon Prevention item Safety CautionsUser Maintenance Usage and MaintenanceEscape Clause PurposePage Names of Components ContentMain Body Components ComponentsControl Panel Components Auto/Manual switchAir check switch Measurement Screen Manual mode, alignment OK Menu Screen Monitor Screen ComponentsMeasurement Screen Auto mode, alignment Contents of Printer Output Standard Accessories Preparations HOW to Install the InstrumentHOW to Connect External I/O Terminals HOW to Connect the Power CableMemo Initial SettingsPreparations Displaying The Menu ScreenTime/Date Setting Setting The Power Save Time RS-232C INPUT/OUTPUT Settings OFF ↔ MODE1 ↔ MODE2 ↔ MODE3 ↔ MODE4 ↔ Setting The Average Value Mode Setting The Buzzer Message Input Auto Setting HOW to SET Printer PaperStep BackPreparations TF50KS-E2C Manual Setting Preparations HOW to Reset from Power Save Status Basic Operations Turn on the PowerPreparations Before Measurement Air CheckSetting the Patient Setting the Safety Stopper ∼10mm Setting the Measurement Mode Measurement Under Auto Mode MemoSetting the Measuring Range AutoAlignment and Measurement Basic Operations Basic Operations Manual mode. Sometimes Auto mode is not available if Display of Measurement Values Measurement Under Manual Mode MemoERR Over ManuBasic Operations Alignment is adjusted correctly. Sometimes the shape Deleting Measurement Values Individual Operations HOW to Print OUT Measurement ValuesHOW to Correct Measurement Values Input via RS-232C INPUT/OUTPUT VIA RS-232COutput via RS-232C Checking Operations Before Requesting ServiceChecking Operations Specifications & Performance Optional AccessoriesReference Automatic instrument table AIT-20 and Table BoardTerminal Pin Arrangement RS-232C Communication SpecificationsConnector Types Transmission Formula Contents of Data TransmissionReference Work ID no Setting RS-232C Communication ConditionsFormat Cleaning the Measuring Window Glass Maintenance and CheckingAccuracy Maintenance Cleaning the Window Glass inside the Nozzle Ordering Consumable Supplies Daily MaintenanceAdjusting the Monitor Screen Replacing the Fuse Paper Jam in PrinterCleaning the Outer Cover Special Notes on CleaningComputerized Tonometer CT-80 Topcon Omni SYSTEMS, INC Computerized Tonometer