If “NG (+)” or “NG 
POWER SWITCH , and check whether or not there is any obstacle in front of the meas- uring nozzle. If there is an object, remove it, and then press the POWER SWITCH .
Press  and perform the checking procedure again.
 and perform the checking procedure again.
If no object is there, a problem has occurred. Turn OFF the POWER SWITCH , unplug the power cable, and call your dealer.
Setting the Patient
| 
 | CAUTION | Never insert your fingers under the measuring head. | 
| 
 | ∗ Inform the patient of this, too. | |
| 
 | 
 | Careless insertion of fingers may cause injury by pinching. | 
| 1 | 
 | Abnormal operation screen | 
| Return to the Measurement screen. | ||
| 2 | Ask the patient to sit in front of the instrument. | |
| 3 | Adjust the automatic instrument table or the chair for height so that the patient can | |
| 
 | put his or her chin on the chinrest in a comfortable position. | |
4 The patient places his or her chin on the chinrest and stops his or her forehead at the forehead rest.
30