Topcon CT-80, COMPUTERIZED TONOMETER instruction manual Memo, HOW to Correct Measurement Values

Page 44

3 Hold the paper and pull it diagonally to cut.

MEMO

To avoid paper jams, cut the paper carefully and evenly.

HOW TO CORRECT MEASUREMENT VALUES

Never set the select switch for more than eight points to avoid MEMO malfunctions.

Though the machine is adjusted for displaying optimal measurement values, the values can be corrected within a -4 +3mmHg range.

1 Make sure the power is OFF.

2 Open the control panel lid.

3 Using a screwdriver, turn the “” of the select switch.

Select switch

Settings: 0: Base setting

For shipment, the switch is set here.

1: +1mmHg

Base setting + 1mmHg

2: +2mmHg

Base setting + 2mmHg

3: +3mmHg

Base setting + 3mmHg

F: -1mmHg

Base setting - 1mmHg

E: -2mmHg

Base setting - 2mmHg

D: -3mmHg

Base setting - 3mmHg

C: -4mmHg

Base setting - 4mmHg

4 Turn the POWER SWITCH ON.

The measurement procedure is exactly the same as with the correction.

43

INDIVIDUAL OPERATIONS

Image 44
Contents CT-80 This instrument features the following IntroductionDisplay for Safe USE Display MeaningSafety Cautions Icons Prevention itemSafety Cautions Icon Prevention itemUsage and Maintenance User MaintenanceEscape Clause PurposePage Content Names of ComponentsComponents Main Body ComponentsControl Panel Components Auto/Manual switchAir check switch Measurement Screen Manual mode, alignment OK Menu Screen Monitor Screen ComponentsMeasurement Screen Auto mode, alignment Contents of Printer Output Standard Accessories HOW to Install the Instrument PreparationsHOW to Connect the Power Cable HOW to Connect External I/O TerminalsInitial Settings MemoPreparations Displaying The Menu ScreenTime/Date Setting Setting The Power Save Time RS-232C INPUT/OUTPUT Settings OFF ↔ MODE1 ↔ MODE2 ↔ MODE3 ↔ MODE4 ↔ Setting The Average Value Mode Setting The Buzzer Message Input HOW to SET Printer Paper Auto SettingStep BackPreparations TF50KS-E2C Manual Setting Preparations HOW to Reset from Power Save Status Turn on the Power Basic OperationsPreparations Before Measurement Air CheckSetting the Patient Setting the Safety Stopper ∼10mm Measurement Under Auto Mode Memo Setting the Measurement ModeSetting the Measuring Range AutoAlignment and Measurement Basic Operations Basic Operations Manual mode. Sometimes Auto mode is not available if Measurement Under Manual Mode Memo Display of Measurement ValuesERR Over ManuBasic Operations Alignment is adjusted correctly. Sometimes the shape Deleting Measurement Values HOW to Print OUT Measurement Values Individual OperationsHOW to Correct Measurement Values Input via RS-232C INPUT/OUTPUT VIA RS-232COutput via RS-232C Checking Operations Before Requesting ServiceChecking Operations Optional Accessories Specifications & PerformanceReference Automatic instrument table AIT-20 and Table BoardTerminal Pin Arrangement RS-232C Communication SpecificationsConnector Types Contents of Data Transmission Transmission FormulaReference Setting RS-232C Communication Conditions Work ID noFormat Cleaning the Measuring Window Glass Maintenance and CheckingAccuracy Maintenance Cleaning the Window Glass inside the Nozzle Ordering Consumable Supplies Daily MaintenanceAdjusting the Monitor Screen Paper Jam in Printer Replacing the FuseSpecial Notes on Cleaning Cleaning the Outer CoverComputerized Tonometer CT-80 Computerized Tonometer Topcon Omni SYSTEMS, INC