MAINTENANCE AND CHECKING

ACCURACY MAINTENANCE

Cleaning the Measuring Window Glass

To secure auto alignment and correct measurement values, clean the measuring window glass after each day’s work.

Clean the glass when “CLEAN THE MEASURING WINDOW GLASS” is displayed on the monitor screen.

To clean the measuring window glass and the window glass inside the measuring nozzle, use ethanol. Using CAUTION other chemicals may cause damage to the patient's eye

during measurement.

 

 

Do not use tissues, as they may make the stain more notice-

 

MEMO

able.

 

Do not use tweezers or gauze, as these may scratch the lens

1

 

and glass surfaces.

Prepare the ethanol.

2

Using a blower, remove dust and dirt from the glass surface.

3

Moisten the applicator with ethanol.

4

Wipe the glass surface lightly with the applicator, from the center outward.

5

Use a new applicator and wipe the glass surface in a similar manner; repeat this

 

several times.

Wiping the glass surface

 

 

To ensure thorough removal of grease from the window glass, be sure to replace the applicator and use a new one for each of these repeated wiping operations.

6 Cleaning is completed when grease is thoroughly removed. If stains cannot be removed easily, call your dealer.

When the measuring window glass becomes stained “CLEAN THE MEASURING WINDOW GLASS” is displayed on the monitor screen.

52

MAINTENANCE AND CHECKING

Image 53
Contents CT-80 Introduction This instrument features the followingDisplay Meaning Display for Safe USEIcons Prevention item Safety CautionsIcon Prevention item Safety CautionsUser Maintenance Usage and MaintenanceEscape Clause PurposePage Names of Components ContentMain Body Components ComponentsControl Panel Components Auto/Manual switchAir check switch Measurement Screen Manual mode, alignment OK Menu Screen Monitor Screen ComponentsMeasurement Screen Auto mode, alignment Contents of Printer Output Standard Accessories Preparations HOW to Install the InstrumentHOW to Connect External I/O Terminals HOW to Connect the Power CableMemo Initial SettingsPreparations Displaying The Menu ScreenTime/Date Setting Setting The Power Save Time RS-232C INPUT/OUTPUT Settings OFF ↔ MODE1 ↔ MODE2 ↔ MODE3 ↔ MODE4 ↔ Setting The Average Value Mode Setting The Buzzer Message Input Auto Setting HOW to SET Printer PaperStep BackPreparations TF50KS-E2C Manual Setting Preparations HOW to Reset from Power Save Status Basic Operations Turn on the PowerPreparations Before Measurement Air CheckSetting the Patient Setting the Safety Stopper ∼10mm Setting the Measurement Mode Measurement Under Auto Mode MemoSetting the Measuring Range AutoAlignment and Measurement Basic Operations Basic Operations Manual mode. Sometimes Auto mode is not available if Display of Measurement Values Measurement Under Manual Mode MemoERR Over ManuBasic Operations Alignment is adjusted correctly. Sometimes the shape Deleting Measurement Values Individual Operations HOW to Print OUT Measurement ValuesHOW to Correct Measurement Values Input via RS-232C INPUT/OUTPUT VIA RS-232COutput via RS-232C Checking Operations Before Requesting ServiceChecking Operations Specifications & Performance Optional AccessoriesReference Automatic instrument table AIT-20 and Table BoardTerminal Pin Arrangement RS-232C Communication SpecificationsConnector Types Transmission Formula Contents of Data TransmissionReference Work ID no Setting RS-232C Communication ConditionsFormat Cleaning the Measuring Window Glass Maintenance and CheckingAccuracy Maintenance Cleaning the Window Glass inside the Nozzle Ordering Consumable Supplies Daily MaintenanceAdjusting the Monitor Screen Replacing the Fuse Paper Jam in PrinterCleaning the Outer Cover Special Notes on CleaningComputerized Tonometer CT-80 Topcon Omni SYSTEMS, INC Computerized Tonometer