Features and Specifications
Specifications
Table 1.2 Specifications for XFR 1200 W Series Supply with GPIB Interface Installed (60 V to 600 V)
Models |
|
|
|
|
|
Program Resolution | 9.2mV | 15.4mV | 23.1mV | 46.2mV | 92.4mV |
Voltage | |||||
Current | 2.8mA | 1.68mA | 1.12mA | 0.56mA | 0.28mA |
OVP | 9.2mV | 15.4mV | 23.1mV | 46.2mV | 92.4mV |
Program Accuracy1 |
|
|
|
|
|
Voltage | 150mV | 150mV | 225mV | 225mV | 250mV |
| ±0.25% | ±0.35% | ±0.35% | ±0.35% | ±0.35% |
Current | 120mA | 80mA | 80mA | 80mA | 50mA |
| ±0.1% | ±0.1% | ±0.1% | ±0.1% | ±0.1% |
OVP | 600mV | 800mV | 1.5V | 3.0V | 6V |
Readback Resolution |
|
|
|
|
|
Voltage | 9.2mV | 15.4mV | 23.1mV | 46.2mV | 92.4mV |
Current | 2.8mA | 1.68mA | 1.12mA | 0.56mA | 0.28mA |
Readback Accuracy1 |
|
|
|
|
|
Voltage | 150mV | 150mV | 225mV | 225mV | 250mV |
| ±0.25% | ±0.35% | ±0.35% | ±0.35% | ±0.35% |
Current | 120mA | 80mA | 80mA | 80mA | 50mA |
| ±0.1% | ±0.1% | ±0.1% | ±0.1% | ±0.1% |
1.Apply accuracy specifications according to the following voltage program accuracy example: Set a model
The expected result will be within the range of 10 volts ± 75mV ± 0.12% of the set voltage of 10 volts.
14 | Operating Manual for GPIB for XHR/XFR Series Power Supply |