Xantrex Technology XHR, XFR, XFR3 manual Calibration

Page 57

Calibration

Voltage Mode Calibration

3.Send command VRLO; IRLO to the power supply. Wait for the supply to settle. Measure and record the output shown on the external voltmeter. Send VRLO again.

4.Send VRHI; IRHI to the supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send VRHI again.

5.Send the command VRDAT <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter after the VRLO and VRHI commands were sent. The processor calculates the offset value required to calibrate the power supply.

When the power supply is calibrated, the low to high voltage readback calibration values (offsets) are stored as constants.

6.Use commands VSET and VOUT? commands to confirm that the calibration was successful and that linearity is observed. Refer to the voltage readback accuracy specification in Section 1.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

Release 3.1

55

Image 57
Contents GPIB-XHR GPIB-XFR GPIB-XFR3 Page Operating Manual for Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyManual Revisions Main SectionsAbout This Manual Contents Contents Description Features and SpecificationsFeatures and Specifications Features and FunctionsFunctions ReadbackSpecifications Models 140 12-100 20-60 35-35 40-30OVP Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Initial Inspection IntroductionInitial Inspection Installation and ConfigurationGpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB Basic Setup Procedure Step # Description Action ReferencePON SRQ Address IEEE-488 Primary Address SelectionPower On Service Request PON SRQ Rear Panel SRQ Switch Position PON SRQ StateSwitch Position Power on results Iset Condition Default Settings Model ExampleIbfind devname Internal PCB Jumper Selections IEEE-488 Controller ConnectionHigh LOW OFF High OFF LOWUser Signals J7 Connector User SignalsJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Gpib Operation OperationExample Ascii Request Service SRQ Set when Bit Decimal Description Reset By Position WeightPower On PON Set when unit LSBCommand Syntax Computer EntryFormat Floating Number Definition Example Description FormIset 2.0A Vset Command Summary AUXB? AUXA?CMODE? DLY?Cmode FAULT?Idata IHICommand Description Command ReferenceAuxa 1 on Auxb 1 onChanging an output setting Response Fold mode where mode is Mask None = Unmask ALL SnspStatus, and Fault Registers on SNSP, ALL, None Accumulated Status, Status, and Fault Registers onResponse Vmax voltage Condition Mnemonic Bit Position Bit Weight Accumulated Status, Status, and Fault Registers11 Error Codes Error Codes LEDs TroubleshootingCalibration Setup Voltage Mode CalibrationVoltage Program Calibration Procedure Readback VoltageCalibration Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc