Xantrex Technology XHR, XFR, XFR3 manual Over Voltage Protection OVP Calibration

Page 60

Calibration

Over Voltage Protection (OVP) Calibration

Over Voltage Protection (OVP) Calibration

We recommend that you perform OVP calibration every six months. Connecting a digital voltmeter as in “Voltage Calibration Setup” is optional.

1.Disconnect all loads from the power supply.

2.Ensure that jumper J65 on the interface PCB is CLOSED to enable remote OVP calibration (Jumper J65 is closed at the factory). See for the location on the PCB of jumper J65.

3.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

4.Send the command OVCAL to the power supply. The ADR LED will light during OVP calibration. Calibration is complete when the ADR LED turns off. This may take a few minutes.

5.Use the OVSET, OVSET?, and VSET commands to trip the OVP level, confirming that the calibration was successful. When you trip the OVP level, the red OVP LED will light and the voltage will drop to zero. Send the command RST to clear the OVP condition. Refer to the OVP program accuracy specification in Section 1.

6.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

58

Operating Manual for GPIB for XHR/XFR Series Power Supply

Image 60
Contents GPIB-XHR GPIB-XFR GPIB-XFR3 Page Operating Manual for How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Contents Features and Specifications DescriptionFeatures and Functions Features and SpecificationsFunctions ReadbackSpecifications Models 140 12-100 20-60 35-35 40-30OVP Models 60-20 100-12 150-8 300-4 600-2 Models 300 12-220 20-130 33-85 40-70 Models 60-46 100-28 150-18 300-9 600-4 Models 130 20-50 33-33 40-25 60-18 Models 100-10 150-7 300-3.5 600-1.7 Introduction Initial InspectionInstallation and Configuration Initial InspectionGpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB Basic Setup Procedure Step # Description Action ReferencePON SRQ IEEE-488 Primary Address Selection AddressPower On Service Request PON SRQ Rear Panel SRQ Switch Position PON SRQ StateSwitch Position Power on results Condition Default Settings Model Example IsetIbfind devname IEEE-488 Controller Connection Internal PCB Jumper SelectionsHigh OFF LOW High LOW OFFUser Signals User Signals J7 ConnectorJ7 User Signal Connector Circuit Block Diagram J7 User Cable with Ferrite Block Operation Gpib OperationExample Ascii Bit Decimal Description Reset By Position Weight Request Service SRQ Set whenPower On PON Set when unit LSBCommand Syntax Computer EntryFormat Description Form Floating Number Definition ExampleIset 2.0A Vset Command Summary AUXA? AUXB?CMODE? DLY?FAULT? CmodeIdata IHICommand Reference Command DescriptionAuxa 1 on Auxb 1 onChanging an output setting Response Fold mode where mode is Snsp Mask None = Unmask ALLStatus, and Fault Registers on Accumulated Status, Status, and Fault Registers on SNSP, ALL, NoneResponse Vmax voltage Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit WeightError Codes 11 Error CodesTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure Readback VoltageCalibration Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc