Sigma C4 manual Delete Program step, Press CLEAR/ENTRY Press CLEAR/ENTRY again

Page 50

6.6Delete Program step

1.Enter the step number to be deleted and press <DISP/CNTL>

2.Press <CLEAR/ENTRY>.

3.Press <CLEAR/ENTRY> again.

The display shows “del nn”, where nn = step number to be deleted.

4, Press <ENTER> to delete the step and move all higher steps down one step.

6.7 Running (Executing) a Program

To start program execution, the beginning program step of the program to be run must be displayed. Displaying a specific program step is covered in Section 6.3.

With the first step of a program displayed, press <START/STOP> to begin program execution at that step. The program will continue to execute until it

reaches program step 100 where it will display end and wait for action from the operator.

Running programs may also be stopped at any time by pressing <START/STOP>. After a program has been stopped in this manner, pressing <START/STOP> again will cause the program to continue executing from the beginning of the step it was executing when it was stopped. Important: If the refrigeration compressor was on when the program execution was interrupted in this way, it will be restarted when program execution is restarted.

At the end of a program, the controller will not turn on further heat or cool, however, the chamber blowers will remain on unless blower shutoff mode has been enabled(see Section 8.5). The blowers can be turned off manually by an operator using the front panel mode switch. Set the switch to OFF to turn off the blowers. The refrigeration compressor (mechanically refrigerated models only) will be turned off when a program completes.

IMPORTANT: If you are controlling an chamber or platform with optional solid state relay power control, including units with compressor control, please read Section 6.8.4 at this time.

CAUTION: Chamber temperatures can exceed 70°C, without heaters energized, if high velocity blowers have been installed. If this presents a problem, be sure the test load is removed promptly at the end of the program run, or enable blower shutoff mode (see Section 8.5) to have blowers not run on at the end of a program execution.

50

C4 Manual Rev 7.5.2

Image 50
Contents Programmable Temperature Controller / Interface Sigma SystemsC4 Manual Rev Table of Contents Intelligent 2 Probe Control C4 Manual Rev C4 Manual Rev C4 Manual Rev C4 Manual Rev Introduction Models C4 & CC-3.5ExplainedGeneral Description Custom Features / Interchangeability Warning Release 7.5.2 FirmwareC4 vs. CC-3 Differences What’s New Firmware Uploads Forced Start from Prom Firmware Internal Error Shutdown Conditions Intelligent 2 Probe Control Probe Averaging Program Mode Step Insert & Delete RSA Remote Mode Setup Parameter Commands Physical Description Front PanelMode Switch Clear Prog Rear Panel Connections J1 TOP200 100c 392 148f General Operation & Error ConditionsC3-5 Rel Startup DisplaysUpdating Firmware Upgrading Firmware by Prom Replacement Making the physical connection for upload Su res Restoring Setup Parameters to Default ValuesRs load Ee busy and then su doneC4 Manual Rev P1f Fahrenheit OperationProbe Out of Range Shutdown System Operating Temperature RangeP1 lo P1 hi P1 err P1 -err All res Internal Error Shutdown ConditionsRes err Su err Software Probe Correction CalibrationStatus and Error Reporting Fail-safe System C4 Manual Rev Intelligent 2 Probe Control How Intelligent 2 Probe Control functions Preparing for Intelligent 2 Probe Control Using Intelligent 2 Probe Control Local Mode Basic Operation Displaying TemperatureDisplaying and Changing the Setpoint S1 102.6 s1fS1 nsp S1f nsp Controlling to a Setpoint Sor -sorSp err C4 Manual Rev Description of a Program Step Program Mode06.0 Ramp time Substep Step Data 00.000.1 00.2Maximizing Ramp Speed & Other Ramp Considerations Clearing Program Memory Reinitializing program steps Displaying Program Steps not during executionPress CLEAR/ENTRY Entering or Changing a Program StepInsert Program Step Running Executing a Program Delete Program stepPress CLEAR/ENTRY Press CLEAR/ENTRY again Program Run Time Information/Considerations 005 sor Sor 007 nop2Loop Special Commands External Compressor On Common Programming Issues C4 Manual Rev IEEE-488 Interface Remote ModeEIA-232 Interface Command Summary by functional group System Information QueriesSystem Information Queries For Celsius mode For parameters numbered 0 through C4 Manual Rev Operation Information Queries & Commands REerror byteCRLF C4 Manual Rev Qccrlf PTnCRLF Setup Parameter Commands SC1 0 2.3 100 WP 6 5 5CRLF Upcrlf System Operation Commands Celsius mode Will hold the current setpoint for 1 hour and 20 minutes U1c 65.0 u1f Error and Status Reporting Overview IEEE-488 GpibTemperature in 20 minutes C4 Manual Rev Error/Status String Bit Definitions C4 Manual Rev Setup Mode PID controller, Integral term Adjustment not Displaying the Field Values Auto-start Mode Setup field Changing the Value of a Setup FieldTwo Probe Mode Setup field Blower Shut-off Mode Setup field Temperature Control Terms PID Setup fields 0, 10, 11Entering probe correction setup data C4 Manual Rev C4 Manual Rev Appendix Programming Examples & Notes00.3 00.000.2 00.4Using shortcuts to shorten program entry time 9900 DISP/CNTLSTART/STOPCLEAR/ENTRY 6 Enter 04.3 04.004.2 04.407.3 07.007.2 07.4C4 Manual Rev Sigma Systems C4 Programming Worksheet Sample Command Structure for IEEE-488 Gpib Operation Disp A$ Installation and Use of TTL Outputs and Input Field Calibration of Model C4 Controller 101102 Troubleshooting 103Noise Immunity 105 Firmware Upload Problems Temperature Control PID Tuning & Problems 107108 109 110 111 U1f 102.7 Technical Support, Repairs & ReturnsIndex 113114 Gpib 115IEEE488 Gpib 117 118 SRQ 119120