9.4 Sample Command Structure for IEEE-488 GPIB Operation
All sample commands are written in BASIC for a Hewlett Packard 85F computer and are representative of the required structure for proper operation of the C4's
Assumptions: C4 address is set to 1.
All options have been implemented.
SI - Select Immediate Mode
OUTPUT 701 "SI"
SP - Select Program Mode
OUTPUT 701 "SP"
ES - Enable the SRQ
OUTPUT 701 "ES"
DS - Disable the SRQ
OUTPUT 701 "DS"
TO - Turn Chamber On
OUTPUT 701 "TO"
TF - Turn Chamber Off
OUTPUT 701 "TF"
CO - Turn Compressor On
OUTPUT 701 "CO"
CF - Turn Compressor Off
OUTPUT 701 "CF"
RS - Request Status Information
OUTPUT 701 "RS"
To receive the status byte, use the following:
Enter 701 using "%,%B";A
DISP A
RE - Request Error Information
OUTPUT 701 "RE"
To receive error byte, use the following:
Enter 701 using "%,%B";A
DISP A