Sigma C4 manual SC1 0 2.3 100

Page 69

Note: Software probe correction is done separately for each probe. The following table shows the U1, C1, U2, C2 setup parameter assignments:

 

Probe 1

Probe 2

U1

F17

F21

C1

F18

F22

U2

F19

F23

C2

F20

F24

The syntax for the SC command is ASCII text as follows:

SCn<space>U1<space>C1<space>U2<space>C2<CR><LF>

where n = number of probe to correct

U1, C1, U2, and C2 are floating point numbers to one decimal place.

Example:

SC1 0 2.3 100 101.3

There is no response to this command from the controller except to set the appropriate bit in the Error Byte in the event that the command or parameters are incorrect.

7.6.2 WP Set PID Constants

A complete discussion of the PID terms entries is in Setup Section 8.6. A discussion of how to use the terms to improve control is in the Appendix in Section 9.8. Read both sections carefully before using this command.

WARNING: Incorrect usage of this command may cause temperature control to be effected with invalid data and may result in unexpected temperature excursions that can cause harm to personnel and damage to equipment.

The WP command writes the P, I, & D data to setup parameter fields F0, F10, & F11 respectively. The data for all 3 parameters used for PID control must be sent with the command. The syntax for the WP command is ASCII text as follows:

WP<space>p<space>i<space>d<CR><LF>

C4 Manual Rev 7.5.2

69

Image 69
Contents Sigma Systems Programmable Temperature Controller / InterfaceC4 Manual Rev Table of Contents Intelligent 2 Probe Control C4 Manual Rev C4 Manual Rev C4 Manual Rev C4 Manual Rev Models C4 & CC-3.5Explained IntroductionGeneral Description Release 7.5.2 Firmware Custom Features / Interchangeability WarningC4 vs. CC-3 Differences What’s New Firmware Uploads Forced Start from Prom Firmware Internal Error Shutdown Conditions Intelligent 2 Probe Control Probe Averaging Program Mode Step Insert & Delete RSA Remote Mode Setup Parameter Commands Front Panel Physical DescriptionMode Switch Clear Prog J1 TOP Rear Panel ConnectionsC3-5 Rel General Operation & Error Conditions200 100c 392 148f Startup DisplaysUpdating Firmware Upgrading Firmware by Prom Replacement Making the physical connection for upload Rs load Restoring Setup Parameters to Default ValuesSu res Ee busy and then su doneC4 Manual Rev Fahrenheit Operation P1fSystem Operating Temperature Range P1 lo P1 hiProbe Out of Range Shutdown P1 err P1 -err Internal Error Shutdown Conditions Res errAll res Software Probe Correction Calibration Su errStatus and Error Reporting Fail-safe System C4 Manual Rev Intelligent 2 Probe Control How Intelligent 2 Probe Control functions Preparing for Intelligent 2 Probe Control Using Intelligent 2 Probe Control Displaying Temperature Local Mode Basic OperationS1 102.6 s1f S1 nsp S1f nspDisplaying and Changing the Setpoint Sor -sor Sp errControlling to a Setpoint C4 Manual Rev Program Mode 06.0Description of a Program Step 00.1 Substep Step Data 00.0Ramp time 00.2Maximizing Ramp Speed & Other Ramp Considerations Displaying Program Steps not during execution Clearing Program Memory Reinitializing program stepsEntering or Changing a Program Step Insert Program StepPress CLEAR/ENTRY Delete Program step Press CLEAR/ENTRY Press CLEAR/ENTRY againRunning Executing a Program Program Run Time Information/Considerations 007 nop2 Loop005 sor Sor Special Commands External Compressor On Common Programming Issues C4 Manual Rev Remote Mode EIA-232 InterfaceIEEE-488 Interface System Information Queries Command Summary by functional groupSystem Information Queries For Celsius mode For parameters numbered 0 through C4 Manual Rev Operation Information Queries & Commands REerror byteCRLF C4 Manual Rev Qccrlf PTnCRLF Setup Parameter Commands SC1 0 2.3 100 WP 6 5 5CRLF Upcrlf System Operation Commands Celsius mode Will hold the current setpoint for 1 hour and 20 minutes U1c 65.0 u1f IEEE-488 Gpib Error and Status Reporting OverviewTemperature in 20 minutes C4 Manual Rev Error/Status String Bit Definitions C4 Manual Rev Setup Mode PID controller, Integral term Adjustment not Displaying the Field Values Changing the Value of a Setup Field Two Probe Mode Setup fieldAuto-start Mode Setup field Temperature Control Terms PID Setup fields 0, 10, 11 Blower Shut-off Mode Setup fieldEntering probe correction setup data C4 Manual Rev C4 Manual Rev Programming Examples & Notes Appendix00.2 00.000.3 00.4Using shortcuts to shorten program entry time DISP/CNTLSTART/STOP 9900CLEAR/ENTRY 6 Enter 04.2 04.004.3 04.407.2 07.007.3 07.4C4 Manual Rev Sigma Systems C4 Programming Worksheet Sample Command Structure for IEEE-488 Gpib Operation Disp A$ Installation and Use of TTL Outputs and Input 101 Field Calibration of Model C4 Controller102 103 TroubleshootingNoise Immunity 105 Firmware Upload Problems 107 Temperature Control PID Tuning & Problems108 109 110 111 Technical Support, Repairs & Returns U1f 102.7113 Index114 115 GpibIEEE488 Gpib 117 118 119 SRQ120