Field | Field Purpose | Permitted Field | Factory Setting | |
# | Values | |||
|
| |||
|
|
|
| |
25 | Chamber/Platform Lower | As required | ||
Operating Temperature Limit | Default = | |||
| ||||
26 | Chamber/Platform Upper | 500/C | As required | |
Operating Temperature Limit | 932/F | Default = 200 | ||
| ||||
27 | Unit Under Test (UUT) Lower | As required | ||
Operating Temperature Limit | Default = | |||
| ||||
28 | Unit Under Test (UUT) Upper | 500/C | As required | |
Operating Temperature Limit | 932/F | Default = 200 | ||
| ||||
| Unit Under Test (UUT) Lower | From 1 to full |
| |
29 | Operating Temperature | controller range | Full controller | |
| Differential Limit | (ie.: 300 for a | range (ie.: 300 | |
| Unit Under Test (UUT) Upper | to +200 | for a | |
30 | Operating Temperature | controller) | +200 controller) | |
| Differential Limit |
|
| |
|
|
|
|
Note: The setup field parameters are stored in
8.1 Displaying the Field Values
The contents of field 0 is displayed upon entering the Setup mode. To display the next field, press <DISP/CNTL>. Pressing <DISP/CNTL> after field 30 will cause field 0 to be