Sigma C4 manual Error/Status String Bit Definitions

Page 79

Error/Status String Bit Definitions

 

Byte 00 - Duplicate of Error Byte

First Release Used

00:0

Aux input TTL high (Door open with optional switch)

. . . . . . . . . . . 7.3.1

00:1

Error in Extended Error String

. . . . . . . . . . . 7.3.1

00:2

Illegal argument string

. . . . . . . . . . . 7.3.1

00:3

Fail-safe tripped

. . . . . . . . . . . 7.3.1

00:4

not used

. . . . . . . . . . . 7.3.1

00:5

Unrecognized command

. . . . . . . . . . . 7.3.1

00:6

Input buffer full

. . . . . . . . . . . 7.3.1

00:7

Time-out error

. . . . . . . . . . . 7.3.1

 

Byte 01 - Duplicate of EIA-232 Status Byte

 

00:0

SRQ enabled

7.5.2

00:1

Aux/Power port ON

7.5.2

00:2

Refer/Compressor port ON

7.5.2

00:3

Not Used

7.5.2

00:4

Setpoint reached

7.5.2

00:5

Not used

7.5.2

00:6

Error

7.5.2

00:7

Interval complete

7.5.2

 

Byte 02 - Setpoint & Probe Errors

 

02:0

Setpoint submitted exceeds upper operating limit 1

7.3.1

02:1

Setpoint submitted exceeds lower operating limit 1

7.3.1

02:2

Probe requested is not defined in F1

7.3.1

 

Byte 03 - Probe Temperature Value Errors

 

03:0

Probe 1 more than 20°C and less than 50°C exceeding high temp operation limit 1

7.3.1

03:1

Probe 1 more than 20°C and less than 50°C exceeding low temp operation limit 1

7.3.1

03:2

Probe 2 more than 20°C and less than 50°C exceeding high temp operation limit 1

7.3.1

03:3

Probe 2 more than 20°C and less than 50°C exceeding low temp operation limit 1

7.3.1

03:4

Probe 1 more than 50°C exceeding high temp operation limit 1

7.3.1

03:5

Probe 1 more than 50°C exceeding low temp operation limit 1

7.3.1

03:6

Probe 2 more than 50°C exceeding high temp operation limit 1

7.3.1

03:7

Probe 2 more than 50°C exceeding low temp operation limit 1

7.3.1

 

Byte 04 - Correction Value Errors

 

04:0

Correction temperatures submitted less than 50/C apart

7.3.1

04:1

Correction factors submitted result in greater than 20/C correction

 

 

at the limits of the controller range

7.3.1

 

Byte 32 - Hardware Errors

 

32:0

EEPROM write failed

7.3.1

32:1

FLASH memory write failed

7.4.9

 

Byte 35 - Port status (set=on, reset=off)

 

35:0

Blower port on

7.3.1

35:1

Main heater port (control heat) on

7.3.1

35:2

Boost heater port on

7.3.1

35:3

Refer (compressor) port on

7.3.1

35:4

Aux output (power) port on

7.3.1

35:5

Coolant port on

7.3.1

35:7

Aux input port on (connected to ground)

7.3.1

C4 Manual Rev 7.5.2

79

Image 79
Contents Sigma Systems Programmable Temperature Controller / InterfaceC4 Manual Rev Table of Contents Intelligent 2 Probe Control C4 Manual Rev C4 Manual Rev C4 Manual Rev C4 Manual Rev Models C4 & CC-3.5Explained IntroductionGeneral Description Release 7.5.2 Firmware Custom Features / Interchangeability WarningC4 vs. CC-3 Differences What’s New Firmware Uploads Forced Start from Prom Firmware Internal Error Shutdown Conditions Intelligent 2 Probe Control Probe Averaging Program Mode Step Insert & Delete RSA Remote Mode Setup Parameter Commands Front Panel Physical DescriptionMode Switch Clear Prog J1 TOP Rear Panel ConnectionsStartup Displays General Operation & Error ConditionsC3-5 Rel 200 100c 392 148fUpdating Firmware Upgrading Firmware by Prom Replacement Making the physical connection for upload Ee busy and then su done Restoring Setup Parameters to Default ValuesRs load Su resC4 Manual Rev Fahrenheit Operation P1fP1 lo P1 hi System Operating Temperature RangeProbe Out of Range Shutdown P1 err P1 -err Res err Internal Error Shutdown ConditionsAll res Software Probe Correction Calibration Su errStatus and Error Reporting Fail-safe System C4 Manual Rev Intelligent 2 Probe Control How Intelligent 2 Probe Control functions Preparing for Intelligent 2 Probe Control Using Intelligent 2 Probe Control Displaying Temperature Local Mode Basic OperationS1 nsp S1f nsp S1 102.6 s1fDisplaying and Changing the Setpoint Sp err Sor -sorControlling to a Setpoint C4 Manual Rev 06.0 Program ModeDescription of a Program Step 00.2 Substep Step Data 00.000.1 Ramp timeMaximizing Ramp Speed & Other Ramp Considerations Displaying Program Steps not during execution Clearing Program Memory Reinitializing program stepsInsert Program Step Entering or Changing a Program StepPress CLEAR/ENTRY Press CLEAR/ENTRY Press CLEAR/ENTRY again Delete Program stepRunning Executing a Program Program Run Time Information/Considerations Loop 007 nop2005 sor Sor Special Commands External Compressor On Common Programming Issues C4 Manual Rev EIA-232 Interface Remote ModeIEEE-488 Interface System Information Queries Command Summary by functional groupSystem Information Queries For Celsius mode For parameters numbered 0 through C4 Manual Rev Operation Information Queries & Commands REerror byteCRLF C4 Manual Rev Qccrlf PTnCRLF Setup Parameter Commands SC1 0 2.3 100 WP 6 5 5CRLF Upcrlf System Operation Commands Celsius mode Will hold the current setpoint for 1 hour and 20 minutes U1c 65.0 u1f IEEE-488 Gpib Error and Status Reporting OverviewTemperature in 20 minutes C4 Manual Rev Error/Status String Bit Definitions C4 Manual Rev Setup Mode PID controller, Integral term Adjustment not Displaying the Field Values Two Probe Mode Setup field Changing the Value of a Setup FieldAuto-start Mode Setup field Temperature Control Terms PID Setup fields 0, 10, 11 Blower Shut-off Mode Setup fieldEntering probe correction setup data C4 Manual Rev C4 Manual Rev Programming Examples & Notes Appendix00.4 00.000.2 00.3Using shortcuts to shorten program entry time DISP/CNTLSTART/STOP 9900CLEAR/ENTRY 6 Enter 04.4 04.004.2 04.307.4 07.007.2 07.3C4 Manual Rev Sigma Systems C4 Programming Worksheet Sample Command Structure for IEEE-488 Gpib Operation Disp A$ Installation and Use of TTL Outputs and Input 101 Field Calibration of Model C4 Controller102 103 TroubleshootingNoise Immunity 105 Firmware Upload Problems 107 Temperature Control PID Tuning & Problems108 109 110 111 Technical Support, Repairs & Returns U1f 102.7113 Index114 115 GpibIEEE488 Gpib 117 118 119 SRQ120