Sigma C4 manual 007 nop2, Loop, sor Sor

Page 52

No Probe 2 Error

If a program step calls for control by probe 2 when setup parameter F1 = 1 (number of probes) the C4 will generate a No Probe 2 Error. Example

007 nop2

You must either remove the step that calls for probe 2 or enable probe 2 with setup parameter F1 (See Section 8) before the program will run. See Section 8.3.

Invalid Loop Count Error

If a program step is a loop start step (substep 4, probe value = 3), then the loop count (substep 0, loop count) must be in the range 1 - 999 integer (no decimal value). If the number is either not an integer, or is outside the range 1 - 999, the

C4 will generate an Invalid Loop Count Error. Example:

003 loop

The invalid loop count must be corrected before the program will run. See Section 6.8.1.

Setpoint Out of Range Error

If a program step requests a setpoint outside the System Operating Range (See Section 3.5), the C4 will generate a Setpoint Out of Range Error. The error message will be preceded by a minus sign (-) if the setpoint requested is lower than the System Operating Range and will not show a minus sign if the requested setpoint is above the System Operating Range. Example:

005 sor

005 -sor

Either the setpoint request or one of the temperature ranges that defines the System Operating Range will have to be changed before the program will run.

52

C4 Manual Rev 7.5.2

Image 52
Contents Programmable Temperature Controller / Interface Sigma SystemsC4 Manual Rev Table of Contents Intelligent 2 Probe Control C4 Manual Rev C4 Manual Rev C4 Manual Rev C4 Manual Rev Introduction Models C4 & CC-3.5ExplainedGeneral Description Custom Features / Interchangeability Warning Release 7.5.2 FirmwareC4 vs. CC-3 Differences What’s New Firmware Uploads Forced Start from Prom Firmware Internal Error Shutdown Conditions Intelligent 2 Probe Control Probe Averaging Program Mode Step Insert & Delete RSA Remote Mode Setup Parameter Commands Physical Description Front PanelMode Switch Clear Prog Rear Panel Connections J1 TOPGeneral Operation & Error Conditions C3-5 Rel200 100c 392 148f Startup DisplaysUpdating Firmware Upgrading Firmware by Prom Replacement Making the physical connection for upload Restoring Setup Parameters to Default Values Rs loadSu res Ee busy and then su doneC4 Manual Rev P1f Fahrenheit OperationP1 lo P1 hi System Operating Temperature RangeProbe Out of Range Shutdown P1 err P1 -err Res err Internal Error Shutdown ConditionsAll res Su err Software Probe Correction CalibrationStatus and Error Reporting Fail-safe System C4 Manual Rev Intelligent 2 Probe Control How Intelligent 2 Probe Control functions Preparing for Intelligent 2 Probe Control Using Intelligent 2 Probe Control Local Mode Basic Operation Displaying TemperatureS1 nsp S1f nsp S1 102.6 s1fDisplaying and Changing the Setpoint Sp err Sor -sorControlling to a Setpoint C4 Manual Rev 06.0 Program ModeDescription of a Program Step Substep Step Data 00.0 00.1Ramp time 00.2Maximizing Ramp Speed & Other Ramp Considerations Clearing Program Memory Reinitializing program steps Displaying Program Steps not during executionInsert Program Step Entering or Changing a Program StepPress CLEAR/ENTRY Press CLEAR/ENTRY Press CLEAR/ENTRY again Delete Program stepRunning Executing a Program Program Run Time Information/Considerations Loop 007 nop2005 sor Sor Special Commands External Compressor On Common Programming Issues C4 Manual Rev EIA-232 Interface Remote ModeIEEE-488 Interface Command Summary by functional group System Information QueriesSystem Information Queries For Celsius mode For parameters numbered 0 through C4 Manual Rev Operation Information Queries & Commands REerror byteCRLF C4 Manual Rev Qccrlf PTnCRLF Setup Parameter Commands SC1 0 2.3 100 WP 6 5 5CRLF Upcrlf System Operation Commands Celsius mode Will hold the current setpoint for 1 hour and 20 minutes U1c 65.0 u1f Error and Status Reporting Overview IEEE-488 GpibTemperature in 20 minutes C4 Manual Rev Error/Status String Bit Definitions C4 Manual Rev Setup Mode PID controller, Integral term Adjustment not Displaying the Field Values Two Probe Mode Setup field Changing the Value of a Setup FieldAuto-start Mode Setup field Blower Shut-off Mode Setup field Temperature Control Terms PID Setup fields 0, 10, 11Entering probe correction setup data C4 Manual Rev C4 Manual Rev Appendix Programming Examples & Notes00.0 00.200.3 00.4Using shortcuts to shorten program entry time 9900 DISP/CNTLSTART/STOPCLEAR/ENTRY 6 Enter 04.0 04.204.3 04.407.0 07.207.3 07.4C4 Manual Rev Sigma Systems C4 Programming Worksheet Sample Command Structure for IEEE-488 Gpib Operation Disp A$ Installation and Use of TTL Outputs and Input Field Calibration of Model C4 Controller 101102 Troubleshooting 103Noise Immunity 105 Firmware Upload Problems Temperature Control PID Tuning & Problems 107108 109 110 111 U1f 102.7 Technical Support, Repairs & ReturnsIndex 113114 Gpib 115IEEE488 Gpib 117 118 SRQ 119120