Sigma C4 manual 04.0, 04.2, 04.3, 04.4, 05.0, 05.2, 05.3, 05.4, 06.2, 06.3, 06.4

Page 94

Step

Displayed

 

 

 

 

Step .

Keys to press

Description of entry

Number

Substep

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

02.0

CLEAR/ENTRY

11.5

ENTER

Go to 11.5/C

 

02.1

CLEAR/ENTRY

15

ENTER

Use 15 minute ramp time

2

02.2

CLEAR/ENTRY

20

ENTER

Hold temperature for 20 minutes

 

02.3

DISP/CNTL

 

Accept default value of 3

 

 

Execute step 3 next

 

02.4

CLEAR/ENTRY

2

ENTER

Use probe 2

 

03.0

CLEAR/ENTRY

-32.3 ENTER

Go to 32.3/C

 

03.1

CLEAR/ENTRY

100

ENTER

Use 1 hour ramp time

3

03.2

CLEAR/ENTRY

100

ENTER

Hold -32.3/C for 1 hour

 

03.3

DISP/CNTL

 

Accept default value of 4

 

03.4

CLEAR/ENTRY 2

ENTER

Use probe 2

 

 

 

04.0

DISP/CNTL

 

Go to 50.1/C

 

04.1

DISP/CNTL

 

Accept default value of 0

4

04.2

CLEAR/ENTRY

9900

ENTER

Do not hold at this temperature

 

04.3

DISP/CNTL

 

Accept default value of 3

 

04.4

CLEAR/ENTRY

5

ENTER

Turn compressor port off

 

05.0

CLEAR/ENTRY

50.0 ENTER

Go to 50/C

 

05.1

DISP/CNTL

 

Accept default value of 0

5

05.2

CLEAR/ENTRY

230

ENTER

Hold 50/C for 2 hours & 30

 

05.3

DISP/CNTL

 

Accept default value of 6

 

 

Execute step 6 next

 

05.4

CLEAR/ENTRY

2

ENTER

Use probe 2

 

06.0

CLEAR/ENTRY 5

ENTER

Set loop count to 5

 

06.1

DISP/CNTL

 

Accept default value

6

06.2

DISP/CNTL

 

Accept default value

 

06.3

DISP/CNTL 1 ENTER

Step number of first step in loop

 

06.4

CLEAR/ENTRY

3

ENTER

Loop command

94

C4 Manual Rev 7.5.2

Image 94
Contents Programmable Temperature Controller / Interface Sigma SystemsC4 Manual Rev Table of Contents Intelligent 2 Probe Control C4 Manual Rev C4 Manual Rev C4 Manual Rev C4 Manual Rev Introduction Models C4 & CC-3.5ExplainedGeneral Description Custom Features / Interchangeability Warning Release 7.5.2 FirmwareC4 vs. CC-3 Differences What’s New Firmware Uploads Forced Start from Prom Firmware Internal Error Shutdown Conditions Intelligent 2 Probe Control Probe Averaging Program Mode Step Insert & Delete RSA Remote Mode Setup Parameter Commands Physical Description Front PanelMode Switch Clear Prog Rear Panel Connections J1 TOP200 100c 392 148f General Operation & Error ConditionsC3-5 Rel Startup DisplaysUpdating Firmware Upgrading Firmware by Prom Replacement Making the physical connection for upload Su res Restoring Setup Parameters to Default ValuesRs load Ee busy and then su doneC4 Manual Rev P1f Fahrenheit OperationP1 lo P1 hi System Operating Temperature RangeProbe Out of Range Shutdown P1 err P1 -err Res err Internal Error Shutdown ConditionsAll res Su err Software Probe Correction CalibrationStatus and Error Reporting Fail-safe System C4 Manual Rev Intelligent 2 Probe Control How Intelligent 2 Probe Control functions Preparing for Intelligent 2 Probe Control Using Intelligent 2 Probe Control Local Mode Basic Operation Displaying TemperatureS1 nsp S1f nsp S1 102.6 s1fDisplaying and Changing the Setpoint Sp err Sor -sorControlling to a Setpoint C4 Manual Rev 06.0 Program ModeDescription of a Program Step Ramp time Substep Step Data 00.000.1 00.2Maximizing Ramp Speed & Other Ramp Considerations Clearing Program Memory Reinitializing program steps Displaying Program Steps not during executionInsert Program Step Entering or Changing a Program StepPress CLEAR/ENTRY Press CLEAR/ENTRY Press CLEAR/ENTRY again Delete Program stepRunning Executing a Program Program Run Time Information/Considerations Loop 007 nop2005 sor Sor Special Commands External Compressor On Common Programming Issues C4 Manual Rev EIA-232 Interface Remote ModeIEEE-488 Interface Command Summary by functional group System Information QueriesSystem Information Queries For Celsius mode For parameters numbered 0 through C4 Manual Rev Operation Information Queries & Commands REerror byteCRLF C4 Manual Rev Qccrlf PTnCRLF Setup Parameter Commands SC1 0 2.3 100 WP 6 5 5CRLF Upcrlf System Operation Commands Celsius mode Will hold the current setpoint for 1 hour and 20 minutes U1c 65.0 u1f Error and Status Reporting Overview IEEE-488 GpibTemperature in 20 minutes C4 Manual Rev Error/Status String Bit Definitions C4 Manual Rev Setup Mode PID controller, Integral term Adjustment not Displaying the Field Values Two Probe Mode Setup field Changing the Value of a Setup FieldAuto-start Mode Setup field Blower Shut-off Mode Setup field Temperature Control Terms PID Setup fields 0, 10, 11Entering probe correction setup data C4 Manual Rev C4 Manual Rev Appendix Programming Examples & Notes00.3 00.000.2 00.4Using shortcuts to shorten program entry time 9900 DISP/CNTLSTART/STOPCLEAR/ENTRY 6 Enter 04.3 04.004.2 04.407.3 07.007.2 07.4C4 Manual Rev Sigma Systems C4 Programming Worksheet Sample Command Structure for IEEE-488 Gpib Operation Disp A$ Installation and Use of TTL Outputs and Input Field Calibration of Model C4 Controller 101102 Troubleshooting 103Noise Immunity 105 Firmware Upload Problems Temperature Control PID Tuning & Problems 107108 109 110 111 U1f 102.7 Technical Support, Repairs & ReturnsIndex 113114 Gpib 115IEEE488 Gpib 117 118 SRQ 119120