Agilent Technologies 6834B, 6814B, 6843A manual Front Panel Operation

Models: 6834B 6843A 6814B

1 79
Download 79 pages 6.57 Kb
Page 48
Image 48
1

4 - Front Panel Operation

1

2

3

Trigger

VOLT T level

VOLT level

VOLT T level

VOLT level

0.0000.0167

VOLT T level

VOLT level

090

 

Figure 4-7. Trigger Delays and Phase Synchronization

 

 

Example

Display

ϕ

This example uses the default trigger parameters. First, access the Voltage

VOLT

120

 

menu and program the immediate and triggered voltage levels, followed by

VOLT:T

150

 

the voltage transient mode.

VOLT:M

STEP

 

Then press Trigger Control and Enter, followed by Shift Trigger.

INIT:IMMED

In this example, you will set a trigger delay. First, access the Voltage menu

VOLT

120

 

and program the immediate and triggered voltage levels, followed by the

VOLT:T

150

 

voltage transient mode.

VOLT:M

STEP

 

Press Trigger Control. Then press until you access the delay parameter.

DELAY

0

 

On the Entry keypad, press . 0 1 6 7 Enter.

DELAY .0167

 

Then press Trigger Control and Enter, followed by Shift Trigger.

INIT:IMMED

This example uses the phase sync mode with no delay, but synchronized at 90.

VOLT

120

 

First, access the Voltage menu and program the immediate and triggered

VOLT:T

150

 

voltage levels, followed by the voltage transient mode.

VOLT:M

STEP

 

Press Trigger Control. Press until you access the delay parameter. If

DELAY

0

 

necessary, set it to 0. Press until you access the sync source command. On

SYNC:SOUR PHASE

 

the Entry keypad, press ° to obtain PHASE. Press Enter.

 

 

Access the Trigger Control menu again and press to access the sync phase reference parameter. On the Entry keypad, program a 90° phase reference by entering 9 0 Enter.

Then press Trigger Control and Enter, followed by Shift Trigger.

SYNC:PHAS 90

INIT:IMMED

48

Page 48
Image 48
Agilent Technologies 6834B, 6814B, 6843A manual Front Panel Operation