Front Panel Operation - 4

 

Example

Display

ϕ

This example uses the immediate slew rate. First, access the Voltage menu

VOLT:M

 

FIXED

 

and press until you access the mode command. On the Entry keypad, press

 

 

 

 

° to obtain FIXED. Press Enter.

 

 

 

 

Access the voltage menu and press until you access the slew command. On

SLEW

50

 

the Entry keypad, press 5 0 Enter to program a slew rate of 50 volts/second.

 

 

 

 

Whenever a new immediate voltage value is entered, the output will slew to

 

 

 

 

the new level at 50 volts/second.

 

 

 

Step mode uses the triggered slew rate. First, access the Voltage menu,

VOLT

 

120

 

program the immediate and triggered voltage levels, and set the slew mode to

VOLT:T

150

 

STEP.

SLEW:M

STEP

 

Access the Voltage menu and press to access the immediate slew

SLEW:

9.9+E37

 

command. On the Entry keypad, enter a value that equals infinity.

 

 

 

 

Access the Voltage menu and press until you access the triggered slew

SLEW:T

50

 

command. On the Entry keypad, enter a value such as 5 0 Enter, which sets

 

 

 

 

the triggered slew rate to 50 volts/second.

 

 

 

 

Then press Trigger Control and Enter, followed by Shift Trigger.

INIT:IMMED

 

After the trigger has been sent, in step mode, the triggered value becomes the

 

 

 

 

new immediate value.

 

 

 

Pulse mode uses the triggered slew rate at the leading edge of the pulse, and

VOLT

 

120

 

the immediate slew rate at the trailing edge of the pulse. First, access the

VOLT:T

150

 

Voltage menu, program the immediate and triggered voltage levels, and set

SLEW:M

 

PULSE

 

the slew mode to PULSE.

 

 

 

 

Access the Pulse menu and program the pulse count, duty cycle, and pulse

COUNT 2

 

period.

DCYCLE

33

 

 

PER

.0166

 

Access the Voltage menu and press to access the immediate slew

SLEW:

9.9+E37

 

command. On the Entry keypad, enter a value that equals infinity.

 

 

 

 

Access the Voltage menu and press until you access the triggered slew

SLEW:T

50

 

command. On the Entry keypad, enter a value such as 5 0 Enter, which sets

 

 

 

 

the triggered slew rate to 50 volts/second.

 

 

 

 

Then press Trigger Control and Enter, followed by Shift Trigger.

INIT:IMMED

When the voltage slew mode is set to LIST, the slew rates are set by the values in the voltage slew list. Refer to the List Transient example for more information on how to program lists. You must program the voltage values and dwell times as explained in that example. You must also program a slew rate for each point in the list (even if it is 9.9+E37).

NOTE: When specifying a dwell time, you must take the slew time into consideration. If the dwell time at any given list point is less than the slew time at the same point, the voltage will never reach its programmed level before the next list point becomes active.