B - Verification and Calibration
Calibrating and Entering Voltage Calibration Values
Action
4.Connect the DVM (ac volts mode) directly to the ac source via the ratio transformer shown in Figure
5.Press Shift Calibration, scroll to CAL VOLT AC, and press Enter.
6.Press Shift Calibration, scroll to CAL LEV P1, and press Enter to select the first calibration point.
7.Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to enter the voltage value displayed on the DVM. (Multiply the DVM readings by the transformer ratio if a ratio transformer is used.)
8.Press Shift Calibration, scroll to CAL LEV P1, use ° to scroll to the P2 parameter, and press Enter. This selects the second calibration point.
9.Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to enter the voltage value displayed on the DVM. (Multiply the DVM readings by the transformer ratio if a ratio transformer is used.)
10.Press Shift Calibration, scroll to CAL LEV P1, use ° to scroll to the P3 parameter, and press Enter.
11.Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to enter the voltage value displayed on the DVM. (Multiply the DVM readings by the transformer ratio if a ratio transformer is used.)
12.Press Shift Calibration, scroll to CAL LEV P1, use ° to scroll to the P4 parameter, and press Enter.
13.Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to enter the voltage value displayed on the DVM. (Multiply the DVM readings by the transformer ratio if a ratio transformer is used.)
The ac source is now holding the new voltage calibration constants in RAM
Calibrating and entering realtime voltage calibration values (Agilent 6843A only)
14.Press Shift Calibration, scroll to CAL VOLT RTIM, and press Enter.
15.Press Shift Calibration, scroll to CAL LEV P1, and press Enter.
16.Press Shift Calibration, scroll to CAL DATA, and use the Entry keypad to enter the voltage value displayed on the DVM. (Multiply the DVM readings by the transformer ratio if a ratio transformer is used.)
Calibrating the OVP trip point
17.Press Shift Calibration, scroll to CAL VOLT PROT, and press Enter.
18.Wait for the ac source to compute the OVP calibration constant. The display returns to Meter mode when the OVP calculation is complete.
The ac source is now holding the new OVP calibration constants in RAM
19.If you are calibrating a
Display
CAL:VOLT:AC
CAL:LEV P1
CAL:DATA 0.00
CAL:LEV P2
CAL:DATA 0.00
CAL:LEV P3
CAL:DATA 0.00
CAL:LEV P4
CAL:DATA 0.00
CAL:VOLT:RTIM
CAL:LEV P1
CAL:DATA 0.00
CAL:VOLT:PROT
64