Agilent Technologies 6843A, 6834B, 6814B manual Verification and Calibration - B, Action

Models: 6834B 6843A 6814B

1 79
Download 79 pages 6.57 Kb
Page 65
Image 65

Verification and Calibration - B

Calibrating and Entering Current Calibration Values

Action

Display

20.Connect the appropriate current shunt and load resistor as shown in figure B- 1. Connect the DVM (ac rms mode) across the current shunt.

If you are calibrating a 3-phase source , make sure that the default phase 1 (φ1) annunciator is lit. Press Phase Select to select a different phase.

21.

Press Shift Calibration, scroll to CAL CURR AC, and press Enter.

CAL:CURR:AC

22.

Press Shift Calibration, scroll to CAL LEV P1, and press Enter.

CAL:LEV P1

23.

Press Shift Calibration and scroll to CAL DATA. Wait for the DVM

CAL:DATA 0.00

 

reading to stabilize. Read the DVM and calculate the first current value

 

 

(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad

 

 

to enter the rms current value.

 

24.

Press Shift Calibration, scroll to CAL LEV P1, use ° to scroll to the P2

CAL:LEV P2

 

parameter, and press Enter.

 

25.

Press Shift Calibration and scroll to CAL DATA. Wait for the DVM

CAL:DATA 0.00

 

reading to stabilize. Read the DVM and calculate the second current value

 

 

(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad

 

 

to enter the rms current value.

 

26.If you are calibrating a 3-phase source , press Phase Select to select the next phase and repeat steps 20 through 25 for phases 2 and 3.

The phase annunciators on the front panel indicate which phase is active.

Calibrating and entering rms current measurement values (Agilent 6843A only)

27.

Press Shift Calibration, scroll to CAL CURR MEAS, and press Enter.

CAL:CURR:MEAS

28.

Press Shift Calibration, scroll to CAL LEV P1, and press Enter.

CAL:LEV P1

29.

Press Shift Calibration and scroll to CAL DATA. Wait for the DVM

CAL:DATA 0.00

 

reading to stabilize. Read the DVM and calculate the first current value

 

 

(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad

 

 

to enter the rms current value.

 

30.

Press Shift Calibration, scroll to CAL LEV P1 command, use ° to scroll

CAL:LEV P2

 

to the P2 parameter, and press Enter.

 

31.

Press Shift Calibration and scroll to the CAL DATA. Wait for the DVM

CAL:DATA 0.00

 

reading to stabilize. Read the DVM and calculate the second current value

 

 

(DVM ac rms voltage divided by the shunt resistance). Use the Entry keypad

 

 

to enter the rms current value.

 

The ac source is now holding the new current calibration constants in RAM

65

Page 65
Image 65
Agilent Technologies 6843A, 6834B, 6814B manual Verification and Calibration - B, Action