Agilent Technologies 6834B Changing the Calibration Password, Saving the Calibration Constants

Models: 6834B 6843A 6814B

1 79
Download 79 pages 6.57 Kb
Page 66
Image 66
Saving the Calibration Constants

B - Verification and Calibration

Calibrating the Output Impedance (Agilent 6843A only)

NOTE: The output voltage and output current must be calibrated before the output impedance can be calibrated.

Action

Display

32.Connect only the output impedance resistor across the output of the ac source. Do not connect any other equipment.

33. Press Shift Calibration, scroll to the CAL IMP command, and press Enter.

CAL:IMP

34.Wait for the ac source to compute the output impedance calibration constant. The display returns to Meter mode when the calculation is complete.

The ac source is now holding the new impedance calibration constants in RAM

Saving the Calibration Constants

CAUTION: Storing calibration constants overwrites the existing ones in non-volatile memory. If you are not sure you want to permanently store the new constants, omit this step. The ac source calibration will then remain unchanged.

 

Action

Display

35.

Press Shift Calibration, scroll to CAL SAVE, and press Enter.

CAL:SAVE

36.

Press Shift Calibration, select CAL OFF, and press Enter to exit Calibration

CAL OFF

 

mode. *RST and *RCL will also set the calibration state to OFF.

 

Changing the Calibration Password

The factory default password is 0. You can change the password when the ac source is in calibration mode (which requires you to enter the existing password). Proceed as follows:

 

Action

Display

1.

Begin by pressing Shift Calibration and scrolling to the CAL ON command.

CAL ON 0.0

2.Enter the existing password from Entry keypad and press Enter

3.

Press Shift Calibration and scroll to the CAL PASS command.

CAL:PASS 0

4.Enter the new password from the keypad. You can use any number with up to six digits and an optional decimal point. If you want the calibration function to operate without requiring any password, change the password to 0 (zero).

NOTE: If you want the calibration function to operate without requiring any password, change the password to 0 (zero).

66

Page 66
Image 66
Agilent Technologies 6834B, 6814B, 6843A manual Changing the Calibration Password, Saving the Calibration Constants