Agilent Technologies 6814B, 6834B, 6843A Calibration Error Messages, Calibration Over the GPIB

Models: 6834B 6843A 6814B

1 79
Download 79 pages 6.57 Kb
Page 67
Image 67
Calibration Error Messages

Verification and Calibration - B

Calibration Error Messages

Errors that can occur during calibration are shown in the following table.

Error

401

402

403

404

405

406

Table B-3. GPIB Calibration Error Messages

Meaning

CAL switch prevents calibration (This is a hardware disable, see the ac source Service Manual.) CAL password is incorrect

CAL not enabled

Computed readback cal constants are incorrect

Computed programming cal constants are incorrect

Incorrect sequence of calibration commands

Calibration Over the GPIB

You can calibrate the ac source by using SCPI commands within your controller programming statements. Be sure you are familiar with calibration from the front panel before you calibrate from a controller. Each front panel calibration command has a corresponding SCPI command.

The SCPI calibration commands are explained in Chapter 3 of the ac source Programming Guide. Calibration error messages that can occur during GPIB calibration are shown in table B-3.

Agilent Calibration Program Listing

Figure B-2 lists the calibration program. This program can be run on any controller operating under Agilent BASIC. The assumed power supply address is 705 and calibration password is 0. If required, change these parameters in the appropriate statements.