Front Panel Calibration Menu

Verification and Calibration - B

Front Panel Calibration Menu

The Entry keypad is used for calibration functions.

Shift

Cal

Press this key to access the calibration menu.

Display

Command Function

CAL ON <value>

Turns calibration mode on when the correct password

 

value is entered.

CAL OFF

Turns calibration mode off

CAL:LEV <char>

Advance to next step in sequence (P1, P2, P3, or P4).

CAL:DATA <value>

Input a calibration measurement.

CAL:VOLT:AC

Begin ac voltage calibration sequence

CAL:VOLT:PROT

Begin voltage protection calibration

CAL:CURR:AC

Begin ac current calibration sequence

CAL:CURR:MEAS

Begin current measurement calibration sequence

CAL:IMP

Begin output impedance calibration sequence

CAL:VOLT:RTIM1

Begin realtime voltage calibration

CAL:SAVE

Saves the calibration constants in non-volatile memory.

CAL:PASS <value>

Set new calibration password.

Notes:

1 Agilent 6843A only

 

value = a numeric value

char = a character string parameter

Use p

and

q

to scroll through the command list.

Use À

and

Ã

to scroll through the parameter list.

Front Panel Calibration

WARNING LETHAL VOLTAGES. Ac sources can supply 425 V peak at their output. DEATH on contact may result if the output terminals or circuits connected to the output are touched when power is applied. These procedures must be performed by a qualified electronics technician or engineer trained on this equipment.

The following procedures assume you understand how to operate front panel keys (see Chapter 4).

Enable Calibration Mode

 

Action

Display

1.

Reset the unit by selecting Shift, Output, and pressing Enter.

*RST

 

If you are calibration a 3-phase source , make sure that the default phase 1

 

 

(φ1) annunciator is lit. Press Phase Select to select a different phase.

 

2.

Press Shift Calibration, scroll to CAL ON, and press Enter.

CAL ON 0.0

3.Enter the calibration password from Entry keypad and press Enter. If the password is correct the Cal annunciator will come on.

If CAL DENIED appears, then an internal switch has been set to prevent the calibration from being changed. (see the Service Manual.)

If the password is incorrect, an error occurs. If the active password is lost, the calibration function can be recovered by setting an internal switch that defeats password protection. (see the Service Manual.)

CAL DENIED

OUT OF RANGE