–171 Invalid expression

–178 Expression data not allowed

Execution Errors –200 through –299 (sets Standard Event Status Register bit #4)

–200 Execution error [generic]

–221 Settings conflict [check current device state]

–222 Data out of range [e.g., too large for this device]

–223 Too much data [out of memory; block, string, or expression too long]

–224 Illegal parameter value [device-specific]

–225 Out of memory

–270 Macro error

–272 Macro execution error

–273 Illegal macro label

–276 Macro recursion error

–277 Macro redefinition not allowed

System Errors –300 through –399 (sets Standard Event Status Register bit #3)

–310 System error [generic]

–350 Too many errors [errors beyond 9 lost due to queue overflow]

Query Errors –400 through –499 (sets Standard Event Status Register bit #2)

–400 Query error [generic]

–410 Query INTERRUPTED [query followed by DAB or GET before response complete]

–420 Query UNTERMINATED [addressed to talk, incomplete programming message received]

–430 Query DEADLOCKED [too many queries in command string]

–440 Query UNTERMINATED [after indefinite response]

Selftest Errors 0 through 99 (sets Standard Event Status Register bit #3)

0No error

1Non-volatile RAM RD0 section checksum failed

2Non-volatile RAM CONFIG section checksum failed

3Non-volatile RAM CAL section checksum failed

4Non-volatile RAM WAVEFORM section checksum failed

5Non-volatile RAM STATE section checksum failed

6Non-volatile RAM LIST section checksum failed

7Non-volatile RAM RST section checksum failed

10 RAM selftest

11 - 31 DAC selftest error, expected <n>, read <reading> Errors 11, 12, 13, 14, 15 apply to DAC12 1A and 1B Errors 16, 17, 18 apply to DAC12 2A

Errors 19, 20, 21 apply to DAC12 2B Errors 22, 23 apply to DAC12 4A Errors 24, 25 apply to DAC12 4B

Errors 26, 27, 28 apply to DAC12 3A and 3B Errors 29, 30, 31 apply to DAC12 5A and 5B

40 Voltage selftest error, output 1