Agilent Technologies 6834B, 6814B, 6843A manual it disables all outputs

Models: 6834B 6843A 6814B

1 79
Download 79 pages 6.57 Kb
Page 54
Image 54
♦it disables all outputs

4 - Front Panel Operation

10 - Switching Between Single- and Three-phase Operation (Agilent 6834B only)

When shipped from the factory, the Agilent 6834B ac source is configured for three-phase operation. You can configure the Agilent 6834B for single-phase operation. This increases the available output power for phase 1 from 1.5 kVA to 4.5 kVA.

NOTE: When you configure the Agilent 6834B for single-phase operation, the phase 2 and phase 3 output terminals are internally connected to the phase 1 output terminal. This distributes the increased available output power across the three output terminals.

Additionally, switching the Agilent 6834B between single-phase and three-phase operation causes the following actions:

it disables all outputs

it reconfigures the current readback and programming calibration constants

it returns all lists and *RCL states to their factory default states

it reboots the unit, after which the newly selected number of phases becomes available

NOTE: This may require you to reprogram the lists and recall states each time the outputs are switched.

Action

1.On the System keypad, press Address.

2.Press to scroll through the address menu until you get to the NOUTPUTS command.

Display

ADDRESS 5

NOUTPUTS 3

3 Use the and ° keys to select the number of output phases; either 1 (single-phase)

NOUTPUTS

3

or 3 (three-phase).

NOUTPUTS

1

54

Page 54
Image 54
Agilent Technologies 6834B, 6814B, 6843A manual it disables all outputs