4 - Front Panel Operation
7 - Measuring Peak Inrush Current
Peak inrush current is a
This example shows you how you can measure the peak inrush current using the front panel meter. The voltage is set to 120 V rms and the output is triggered at an output phase of 75°, which optimizes the conditions under which inrush current is applied to the unit under test.
| Action | Display | |
1. | Set the immediate voltage to 0. Press Voltage, then press 0 and Enter. | VOLT | 0 |
2. | Set the triggered voltage to 120 V rms. In the Voltage menu, press to | VOLT:T | 120 |
| access the triggered voltage command. Then press 1 2 0 Enter. |
|
|
3. | Set the voltage mode to step. In the Voltage menu, press to access the | VOLT:M | STEP |
| mode command. Press ° to scroll to STEP and press Enter. |
|
|
4. | Make sure the unit is operating in the low voltage range. (The low voltage | RANGE | 150 |
| range has twice the output current rating of the high voltage range.) In the |
|
|
| Voltage menu, press to access the range command. If necessary, press ° to |
|
|
| set the range to 150 and press Enter. |
|
|
5. | Make sure the triggered voltage slew rate is set to the fastest possible speed. | SLEW:T 9.9000+E37 | |
| In the Voltage menu, access the triggered slew command. If necessary, reset |
|
|
| the slew rate to a faster speed. |
|
|
6. | Make sure the peak current and rms current limits are set to high values. In the | CURR:LEV 6.5 | |
| Current menu, access the rms current limit and then the peak current limit | CURR:PEAK 40 | |
| commands. If necessary, reset the rms current and the peak current limits to |
|
|
| higher values. |
|
|
| (Agilent 6811B units have an rms current limit of 3.25 A. Agilent 6813B units |
|
|
| have an rms current limit of 13 A and a peak current limit of 80 A) |
|
|
7.Synchronize the trigger source with a reference phase angle. In the Trigger Control menu, press to access the sync source command. Press ° to obtain PHASE and press Enter.
8.Set the reference phase angle to 75°. In the Trigger Control menu, press to access the sync phase command. Then press 7 5 Enter.
SYNC:SOUR PHASE
SYNC:PHAS 75
9. | Initiate (or enable) the unit for one immediate trigger from the front panel. | INIT:IMMED | |
| Press Trigger Control and Enter. |
|
|
10. | Set the meter function to measure nonrepetitive peak inrush current. In the | 0 A | PK NR |
| Meter menu, press to access the peak inrush current display. |
|
|
11. | Enable the output by pressing Output On/Off. | 0 V | 60 HZ |
12. | Send the trigger to step the output from 0 V to 120 V. Press Shift Trigger. | 48 A | PK NR |
| The inrush current is displayed on the Meter. |
|
|
On
52