Agilent Technologies 6814B, 6834B, 6843A manual Front Panel Operation, Display, Command Function

Models: 6834B 6843A 6814B

1 79
Download 79 pages 6.57 Kb
Page 37
Image 37
Manual background

Shift

Phase

Front Panel Operation - 4

Press this key to access the phase menu list.

Display

Command Function

PHASE <value>

Set immediate output phase (φ)

PHASE:T <value>

Set triggered output phase (φ)

PHASE:M <char>

Select the phase mode (φ) (FIXED, STEP, PULSE or LIST)

Shape

Press this key to access the shape menu list.

Display

Command Function

SHAPE <char>

Select the immediate output wave shape 1 (SINE, SQUARE,

 

or CSINE) CSIN = clipped sine wave

SHAPE:T <char>

Select the triggered output wave shape (SINE, SQUARE, or

 

CSINE) CSIN = clipped sine wave

SHAPE:M <char>

Select the shape mode (FIXED, STEP, PULSE or LIST)

CLIP <value>

Set the clipping level of the CSIN wave shape. This

 

specifies the point where clipping starts as a percentage of

 

the peak amplitude or percentage of THD.

Pulse

Press this key to access the pulse menu list.

Display

Command Function

WIDTH <value>

Set the pulse width

COUNT <value>

Set the number of output pulses

DCYCLE <value>

Set the pulse duty cycle as a percentage of the pulse period

PER <value>

Set the pulse period

HOLD <char>

Set the parameter that will be held constant as the other

 

parameters change (WIDTH or DCYCLE)

Shift

Output

Press this key to access the output menu list.

Display

Command Function

*RST

Execute *RST command to place the ac source in the

 

factory-default state

TTLT:SOUR <char>

Select Trigger Out source coupling (BOT, EOT or LIST)

 

BOT = beginning of transient

 

EOT = end of transient

 

LIST = TTLT trigger (see Programming Guide)

TTLT:STATE <value>

Set Trigger Out state (ON or OFF)

PON:STATE <char>

Select power-on state command (RST or RCL0)

RI <char>

Sets remote inhibit mode (LATCHING, LIVE, or OFF)

DFI <char>

Sets discrete fault indicator state (ON or OFF)

DFI:SOUR <char>

Select the DFI source (QUES, OPER, ESB, RQS, or OFF)

 

(see Chapter 4 of Programming Guide)

Notes:

1 User defined waveshapes will also appear in this list when created. φ = phase selectable on Agilent 6834B

value = a numeric value

char = a character string parameter

p

and

q

scroll through the command list.

Ø

￿

 

and

 

scroll through the parameter list.

37