Language Dictionary - 8
117
[SOURce:]VOLTage
This command sets the output voltage level of the dc source.
Command Syntax [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]<NRf+>
Parameters see Table 8-3
Default Suffix V (volts)
*RST Value 0
Examples VOLT 2.5 !set output voltage to 2.5V
Query Syntax [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters <NR3>
Related Commands VOLT:TRIG VOLT:PROT
[SOURce:]VOLTage2
Agilent 66319B/66319D only
This command sets the output voltage level of the auxiliary output.
Command Syntax [SOURce:]VOLTage2[:LEVel][:IMMediate][:AMPLitude]<NRf+>
Parameters see Table A-3
Default Suffix V (volts)
*RST Value 0
Examples VOLT2 500mV !set output2 voltage to 0.5V
Query Syntax [SOURce:]VOLTage2[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters <NR3>
Related Commands VOLT2:TRIG
[SOURce:]VOLTage:PROTection
This command lets you limit the maximum allowable output voltage that can be programmed either from
the front panel or over the GPIB. This feature is in addition to the automatic overvoltage protection
circuit, and is useful in situations where accidentally programming higher output voltages within the
operating range of the dc source can permanently damage the phone under test.
NOTE: This command does not program the tracking OVP circuit, which automatically
tracks the output voltage and trips when the output voltage exceeds the programmed
voltage by two volts. Also, the programmable voltage protection cannot be disabled by
VOLTage:PROTection:STATe.
Command Syntax [SOURce:]VOLTage:PROTection[:LEVel] <NRf+>
Parameters see Table 8-3
Default Suffix V (volts)
*RST Value 22 V
Examples VOLT:PROT 10 !set voltage limit to 10V
Query Syntax [SOURce:]VOLTage:PROTection[:LEVel]?
Returned Parameters <NR3>
Related Commands VOLT VOLT:TRIG