Specifications - A
141
Table A-2. Supplemental Characteristics (continued)
Parameter Agilent 66321B/D;
Agilent 66319B/D
output 1 only
Agilent 66319B/D
output 2 only
Typical Transient Voltage
Undershoot Values
(actual values are dependent on the
test setup)
With short load leads:
(< 1 meter)
With long load leads:
(up to 6 meters)
30mV w/6µF load cap
25mV w/20µF load cap
40mV w/6µF load cap
30mV w/20µF load cap
NA
Dynamic Measurement System 1 Buffer Length:
Sample Rate Range: 1 4096 points
15.6µs 31200s NA
Measurement Time
(includes 30 ms2 acquisition time
and 20 ms processing overhead)
voltage or current 50 ms average
Command Processing Time
(time for output to begin to change
following receipt of digital data)
4 ms average
Savable Instrument States 4 (in locations 0 to 3)
GPIB Interface Capabilities Language:
Interface: SCPI
AH1, C0, DC1, DT1, E1,
L4, PP0, RL1, SH1, SR1, T6
FLT Output Characteristics
INH/Trigger Characteristics
Maximum ratings:
FLT Output Terminals:
INH/Trigger Terminals:
16.5 Vdc between terminals 1 and 2; 3 and 4;
and from terminals 1 or 2 to chassis ground
Low-level output current = 1.25 mA max.
Low-level output voltage = 0.5 V max.
Low-level input voltage = 0.8 V max.
High-level input voltage = 2 V min.
Low-level input current = 1 mA
Pulse width = 100 µs minimum
INH time delay = 4 ms typical
Trigger latency = 15.6 µs to + 32 µs
Digital I/O Characteristics Maximum ratings:
Digital OUT Port 0,1,2
(open collector)
Digital IN Port 2:
(internal pull-up)
same as FLT/INH/Trigger Characteristics
Output leakage @ 16V = 0.1 mA (ports 0,1)
= 12.5 mA (port 2)
Output leakage @ 5V = 0.1 mA (ports 0,1)
= 0.25 mA (port 2)
Low-level output sink current @ 0.5 V = 4 mA
Low-level output sink current @ 1 V = 50 mA
Low-level input current @ 0.4 V = 1.25 mA
High-level input current @ 5 V = 0.25 mA
Low-level input voltage = 0.8 V max.
High level input voltage = 2.0 V min.
Isolation to Ground
Maximum from either
output terminal to chassis: 50 Vdc
Mains Input Ratings
(at full load from 47–63 Hz)
100 Vac (87-106 Vac):
120 Vac (104-127 Vac):
220 Vac (191-233 Vac):
230 Vac (207-253 Vac):
66321B/D
1.6 A, 125 W
1.4 A, 125 W
0.8 A, 125 W
0.75A, 125 W
66319B/D
2.2 A, 170 W
1.7 A, 170 W
0.96 A, 170 W
0.85A, 170 W
1For a pulse waveform, the accuracy of any individual data point in the buffer depends on the rise time of the pulse. Fo r a cu rrent
pulse of 1.4A with a rise time constant of 50µs, the error in measurement of a single data point during the rise time is 10mA.
2May be reduced by changing the default conditions of 2048 data points but measurement uncertainty due to noise will increase.