Language Dictionary - 8
131
Command Syntax TRIGger:SEQuence2:SLOPe:DVM<slope>
TRIGger:ACQuire:SLOPe:DVM<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:DVM POS TRIG:ACQ:SLOP:DVM EITH
Query Syntax TRIGger:SEQuence2:SLOPe:DVM?
TRIGger:ACQuire:SLOPe:DVM?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:SLOPe:VOLTage TRIGger:ACQuire:SLOPe:VOLTage
This command sets the slope of an internally triggered voltage measurement.
POSitive triggering occurs on the rising edge.
NEGative triggering occurs on the falling edge.
EITHer triggering occurs on either edge.
Command Syntax TRIGger:SEQuence2:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:VOLT POS TRIG:ACQ:SLOP:VOLT EITH
Query Syntax TRIGger:SEQuence2:SLOPe:VOLTage?
TRIGger:ACQuire:SLOPe:VOLTage?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:SLOP:CURR
TRIGger:SEQuence2:SOURce TRIGger:ACQuire:SOURce
These commands select the trigger source for measurement triggers as follows:
BUS GPIB device, *TRG, or <GET> (Group Execute Trigger)
INTernal trigger is generated internally when the measured waveform crosses the trigger
level with the selected slope.
EXTernal Selects the external trigger input on the 4-pin control connector as the trigger
source (only applies to units with firmware revision A.03.01 and up)
Command Syntax TRIGger:SEQuence2:SOURce<source>
TRIGger:ACQuire:SOURce<source>
Parameters BUS | INTernal
*RST Value INTernal
Examples TRIG:ACQ:SOUR BUS
Query Syntax TRIGger:SEQuence2:SOURce?
TRIGger:ACQuire:SOURce?
Returned Parameters <CRD>