41 
4 
Turn-On Checkout Checkout Procedure
Successful tests in this chapter provide a high degree of confidence that your unit is operating properly. 
For performance tests, see appendix B. 
NOTE:  To perform the checkout procedure, you will need a wire for shorting the output 
terminals together. 
The following procedure assumes that the unit turns on in the factory-default state. If you need more 
information about the factory default state, refer to the *RST command in chapter 8. Note that the values 
shown in the Display column may not exactly match the values that appear on the front panel of the unit. 
If you have not already done so, connect the power cord to the unit and plug it in. Connect the output 
connector to the back of the unit with the sense jumpers installed. 
  Procedure   Display   Explanation 
1.  Turn the unit on. The dc 
source undergoes a self-test 
when you first turn it on. 
********** 
ADDRESS 5 
0.000V     0.0001A 
During selftest, all display segments are briefly lit, 
followed by the GPIB Address.  
The display then goes into meter mode with the Dis 
annunciator on, and all others off. In Meter mod e  the 
n.nnnV digits indicate the output voltage and the 
.nnnnA digits indicate the output current. The 
flashing digit on the display indicates the digit that 
will be affected if changes are made to the displayed 
values using the rotary control or the # and & keys.  
You will only see the changes if the output is ON.  
NOTE:  Press the Meter key to exit a menu at any time and return to meter mode. If the Err 
annunciator on the display is on, press the Shift key followed by the Error key to see the 
error number. See table 4-1 at the end of this chapter. 
2.  Check that the fan is on.     You should be able to hear the fan and feel the air 
coming from the back of the unit. 
3.  Unplug the output connector 
from the back of the unit.  -0.224V    0.0000A  The output voltage indicates approximately -0.2 volts 
because the output sense connections have opened. 
4  P ress Output, scroll to 
SENSE:PROT and select 
ON. Press Enter 
SENSE:PROT ON  Enables the open sense detect circuit. 
5.  Press Output On/Off -0.224V   0.0000A  The open sense detect circuit disables the output. The 
Dis annunciator is off, but the Prot annunciator is on. 
6.  Press Protect +/- SENSE OPEN  Display indicates the protection condition.