Language Dictionary - 8
129
TRIGger:SEQuence2:LEVel:CURRent TRIGger:ACQuire:LEVel:CURRent
This command sets the trigger level for internally triggered current measurements. A positive current
trigger occurs when the current level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative current trigger occurs when the
current level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:LEVel:CURRent<NRf+>
TRIGger:ACQuire:LEVel:CURRent<NRf+>
Parameters 0 to MAX (see table 8-3)
Unit A (amperes)
*RST Value 0
Examples TRIG:SEQ2:LEV:CURR 5 TRIG:ACQ:LEV:CURR MAX
TRIG:ACQ:LEV 2
Query Syntax TRIGger:SEQuence2:LEVel:CURRent?
TRIGger:ACQuire:LEVel:CURRent?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:LEV:VOLT TRIG:SEQ2:HYST:CURR
TRIGger:SEQuence2:LEVel:DVM TRIGger:ACQuire:LEVel:DVM
Agilent 66321D/66319D only
This command sets the trigger level for DVM measurements. A positive trigger occurs when the input
signal changes from a value less than the lower hysteresis band limit to a value greater than the upper
hysteresis band limit. Similarly, a negative trigger occurs when the input signal changes from a value
greater than the upper hysteresis band limit to a value less than the lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:LEVel:DVM<NRf+>
TRIGger:ACQuire:LEVel:DVM<NRf+>
Parameters 0 to MAX (see table A-3)
Unit V (volts)
*RST Value 0
Examples TRIG:SEQ2:LEV:DVM 5 TRIG:ACQ:LEV:DVM MAX
Query Syntax TRIGger:SEQuence2:LEVel:DVM?
TRIGger:ACQuire:LEVel:DVM?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:HYST:DVM