8 – Language Dictionary
128
TRIGger:SEQuence2:HYSTeresis:DVM TRIGger:ACQuire:HYSTeresis:DVM
Agilent 66321D/66319D only
This command defines a band around the trigger level through which the input signal must pass before a
DVM measurement can occur. The band limit above and below the trigger level is one half of the
hysteresis value added to or subtracted from the trigger level. For a positive trigger to occur, the
excursion of an input signal in the positive direction must start below the lower hysteresis band limit and
pass through the upper hysteresis band limit. For a negative trigger to occur, the excursion of an input
signal in the negative direction must start above the upper hysteresis band limit and pass through the
lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:HYSTeresis:DVM<NRf+>
TRIGger:ACQuire:HYSTeresis:DVM<NRf+>
Parameters 0 to MAX (see table A-3)
Unit V (volts)
*RST Value 0
Examples TRIG:SEQ2:HYST:DVM 0.5 TRIG:ACQ:HYST:DVM 0.5
Query Syntax TRIGger:SEQuence2:HYSTeresis:DVM?
TRIGger:ACQuire:HYSTeresis:DVM?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:HYSTeresis:VOLTage TRIGger:ACQuire:HYSTeresis:VOLTage
This command defines a band around the trigger level through which the signal must pass before an
internal measurement can occur. The band limit above and below the trigger level is one half of the
hysteresis value added to or subtracted from the trigger level.
For a positive trigger to occur, the excursion of an output waveform in the positive direction must start
below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative
trigger to occur, the excursion of an output waveform in the negative direction must start above the upper
hysteresis band limit and pass through the lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:HYSTeresis:VOLTage<NRf+>
TRIGger:ACQuire:HYSTeresis:VOLTage<NRf+>
Parameters 0 to MAX (see table 8-3)
Unit V (volts)
*RST Value 0
Examples TRIG:SEQ2:HYST:VOLT 2
TRIG:ACQ:HYST:VOLT 2
Query Syntax TRIGger:SEQuence2:HYSTeresis:VOLTage?
TRIGger:ACQuire:HYSTeresis:VOLTage?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:HYST:CURR TRIG:SEQ2:LEV:VOLT