8 – Language Dictionary
130
TRIGger:SEQuence2:LEVel:VOLTage TRIGger:ACQuire:LEVel:VOLTage
This command sets the trigger level for internally triggered voltage measurements. A positive voltage
trigger occurs when the voltage level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative voltage trigger occurs when the
voltage level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax TRIGger:SEQuence2:LEVel:VOLTage<NRf+>
TRIGger:ACQuire:LEVel:VOLTage<NRf+>
Parameters 0 to MAX (see table 8-3)
Unit V (volts)
*RST Value 0
Examples TRIG:SEQ2:LEV:VOLT 5
TRIG:ACQ:LEV:VOLT MAX
Query Syntax TRIGger:SEQuence2:LEVel:VOLTage?
TRIGger:ACQuire:LEVel:VOLTage?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2:LEV:CURR TRIG:SEQ2:HYST:VOLT
TRIGger:SEQuence2:SLOPe:CURRent TRIGger:ACQuire:SLOPe:CURRent
This command sets the slope of an internally triggered current measurement.
POSitive triggering occurs on the rising edge.
NEGative triggering occurs on the falling edge.
EITHer triggering occurs on either edge.
Command Syntax TRIGger:SEQuence2:SLOPe:CURRent<slope>
TRIGger:ACQuire:SLOPe:CURRent<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:CURR POS TRIG:ACQ:SLOP:CURR EITH
Query Syntax TRIGger:SEQuence2:SLOPe:CURRent?
TRIGger:ACQuire:SLOPe:CURRent?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:SLOP:VOLT
TRIGger:SEQuence2:SLOPe:DVM TRIGger:ACQuire:SLOPe:DVM
Agilent 66321D/66319D only
This command sets the slope of the DVM input signal.
POSitive measurement triggering occurs on the rising edge.
NEGative measurement triggering occurs on the falling edge.
EITHer measurement triggering occurs on either edge.