Appendix C

RF Switch Error Messages

The following error messages are unique to the E1470A. See the appropriate command module or VXI Controller module manual for a more complete list of possible error messages.

Error

Message Generated

Description

Commands that

Number

 

 

may cause error

 

 

 

 

-108

“Parameter not allowed”

A parameter was specified that is not valid for the

 

 

 

command. More than 80 channels specified in

 

 

 

command.

DIAG:OPEN;CLOS;DI:CLOS?;IAG:OPEN?

 

 

 

 

-222

“Data out of Range”

Invalid numerical state parameter.

 

 

 

 

 

-240

“Hardware Error”

More than 60 msec was required for the relays to

SAVRCL

 

 

settle.

PC

 

 

 

 

2001

“Invalid channel number”

Invalid channel number in <source> parameter

*WAI?

 

 

 

 

2022

“Invalid relay number”

Invalid relay number in command

 

 

 

 

 

2023

“Invalid common bank

Invalid <comm> parameter in command.

DIAG:OPEN;CLOS;DI:CLOS?;IAG:OPEN?

 

number”

 

 

 

 

 

 

2024

“Invalid source bank

Invalid source number in <source> parameter

 

 

number”

 

 

 

 

 

 

2025

“Invalid Common-Source

Even though the <comm> and <source>

 

 

Combination”

parameters are valid, the combination is not valid.

 

 

 

The specified source cannot connect to the

 

 

 

specified COM terminal.

[ROUT:]PATH[:COMM]?

 

 

 

 

Appendix C

RF Switch Error Messages 53

Page 53
Image 53
Agilent Technologies E1470A user manual Appendix C RF Switch Error Messages