Agilent Technologies E1470A user manual Appendix a RF Switch Specifications

Models: E1470A

1 58
Download 58 pages 63.08 Kb
Page 39
Image 39

Appendix A

RF Switch Specifications

Configuration:

 

3dB Bandwidth:

 

80 signal connections

 

3:1 Multiplexer:

500 MHz

60 inputs (channel numbers xx0 through xx2)

30:1 Multiplexer:

200 MHz (30:1 specifications apply

20 commons (channel numbers COMxx)

 

 

for channels 000 - 132 to COM05 or

One 60:1, two 30:1,... up to 20 3:1 multiplexers

 

 

channels 200 - 332 to COM25

can be configured

 

 

 

multiplexers)

 

 

 

60:1 Multiplexer:

100 MHz

 

 

 

Terminated Isolation:

 

VSWR (Voltage Standing Wave Ratio) for a 3:1

10 MHz:

80 dB

 

Multiplexer:

 

 

100 MHz:

60 dB

 

100 MHz:

1.4

 

200 MHz:

50 dB

 

200 MHz:

1.45

 

500 MHz:

40 dB

 

500 MHz:

1.7

 

 

 

VSWR (Voltage Standing Wave Ratio) for a 30:1

VSWR (Voltage Standing Wave Ratio) for a 60:1

Multiplexer:

 

 

Multiplexer:

 

 

200 MHz:

1.5

 

100 MHz:

1.5

 

(30:1 muxs are channels 000 - 132 to COM05 or

 

 

 

channels 200 - 332 to COM25 multiplexers)

 

 

 

 

 

 

Characteristic Impedance:

 

Power Consumption:

50

 

 

+5 Volt power supply, 3.5A (all relays closed)

 

 

 

Relay Ratings:

 

A16 Register-based device:

Maximum RF Power:

10W

A16 Register-based Slave-only device as defined in the

Switch Life: no load:

5x106 closures

VMEbus standard and the VXIbus specification, rev. 1.4.

0.01A @ 24 Vdc:

3x105 closures

 

 

 

10 Watts RF:

1x105 closures

 

 

 

Power-on/Reset State:

 

 

 

 

Channel xx0 connected to COMxx. Channel 0 in each 3:1

 

 

 

is connected to its respective common.

 

 

 

 

 

 

 

 

 

Appendix A

RF Switch Specifications 39

Page 39
Image 39