Agilent Technologies E3615A, E3614A, E3617A, E3616A service manual Current Monitoring

Models: E3615A E3614A E3616A E3617A

1 47
Download 47 pages 24.48 Kb
Page 43
Image 43

 

 

Figure 6

 

Set the CC switch down, and all others up.

 

Iin = 1/A x Iout

 

 

 

 

Iout = A x Vin

 

 

 

 

Where

Iout is the power supply output current in amps.

 

Iin is the programming current in μamps.

 

A is the gain.

 

 

 

 

 

 

 

Model

 

A (A/μA)

1/A (μV/A)

 

 

 

 

 

 

E3614A

 

0.055

18

 

 

 

 

 

 

E3515A

 

0.0278

35.9

 

 

 

 

 

 

E3616A

 

0.0158

63.4

 

 

 

 

 

 

E3617A

 

0.00928

108

 

 

 

 

 

Programming currents can be increased by adding a resistor across the CC+ and CC-. A 10 volts drop across R1 represents full scale current of the power supply. When a 1 kohm resistor is added across R1, the programming currents are as follows with the programming current in mA.

Model

A (A/mV)

1/A (mA/A)

Parallel resistor required for a

1 amp/mamp value of A (kohm)

 

 

 

 

 

 

 

E3614A

0.594

1.69

1.7

 

 

 

 

E3515A

0.297

3.37

3.45

 

 

 

 

E3616A

0.168

5.95

6.28

 

 

 

 

E3617A

0.0989

1.01

11.2

 

 

 

 

Current Monitoring

Current of the power supply can be monitored across the internal current monitoring resistor. One side of the resistor is at the +output and A3; the other side of the resistor is at A1. The table below shows the resistor value and conversion factors. To obtain the current divide the measured voltage by the resistor value or multiply the amps/V times the voltage measured.

 

Model

Resistor

amps/V

 

 

value (Ω)

 

 

 

 

 

 

 

E3614A

0.1

10

 

 

 

 

 

 

 

E3515A

0.2

5

 

 

 

 

 

 

 

E3616A

0.6

1.67

 

 

 

 

 

 

 

E3617A

0.89

1.12

 

 

 

 

 

 

 

 

 

 

 