a.Perform turn-on checkout procedure given in page 1-7.

b.Perform the CV and CC Load Regulation perfor- mance tests given in the following paragraphs respectively.

PERFORMANCE TESTS

The following paragraphs provide test procedures for verify- ing the power supply's compliance with the specifications of Table 1. Please refer to adjustment and calibration or trouble- shooting procedure if you observe any out of specification performance.

Measurement Techniques

Setup for All Tests. Measure the output voltage directly at the +S and -S terminals on the rear panel; in this way the monitoring device sees the same performance as the feedback amplifier within the power supply. Failure to connect the monitoring device to the proper points shown in Figure A-3 will result in the mea- surement not of the power supply characteristics, but of the power supply plus the resistance of the leads between its output terminals and the point of connection.

Use separate leads to all measuring devices to avoid the sub- tle mutual coupling effects that may occur between measur- ing devices unless all are returned to the low impedance terminals of the power supply. Twisted pairs or shielded cable should be used to avoid pickup on the measuring leads.

Electronic Load. The test and calibration procedures use an electronic load to test the supply quickly and accurately. An electronic load is considerably easier to use than load resis- tor. It eliminates the need for connecting resistors or rheostats in parallel to handle the power, it is much more stable than carbon-pile load, and it makes easy work of switching between load conditions as is required for the load regulation and load transient response tests.

Current Monitoring Resistor Rs. To eliminate output-current measurement error caused by voltage drops in the leads and connections, connect the current monitoring (sampling) resis- tor between -OUT and the load as a four-terminal device. Fig- ure A-3 shows correct connections. Connect the current monitoring test leads inside the load lead connections directly at the monitoring resistor element. Select a resistor with sta- ble characteristics and lower temperature coefficient (see Table A-1).

Figure A-3. Current Monitoring Resistor Connections

Table A-1. Test Equipment Required

TYPE

 

REQUIRED CHARACTERISTICS

USE

RECOMMENDED MODEL

 

 

 

 

 

Oscilloscope

 

Sensitivity : 1 mV

P, T

Agilent 54600A

 

 

Bandwidth : 20 MHz/100 MHz

 

 

 

 

Input : Differential, 50 ohm and 100 ohm

 

 

 

 

 

 

 

RMS Voltmeter

 

True rms, 20 MHz bandwidth

P

 

 

 

Sensitivity : 1 mV

 

 

 

 

Accuracy : 5%

 

 

 

 

 

 

 

Multimeter

 

Resolution : 100 nV

P, A, T

Agilent 34401A

 

 

Accuracy : 0.0035%

 

 

 

 

 

 

 

Electronic Load

 

Voltage Range : 240 Vdc

P

Agilent 6063A

 

 

Current Range : 10 Adc

 

 

 

 

Open and short switches

 

 

 

 

Transient on/off

 

 

 

 

 

 

 

Load Resistor (RL)

 

1.3 ohm 60 W, 6.6 ohm 60 W, 20.5 ohm 60 W,

P

 

 

 

60 ohm 60 W

 

 

 

 

 

 

 

Current Monitoring

 

0.1 ohm 0.1% 10 W, 1 ohm 1% 5 W

P, A

 

(Sampling) Resistor (RS)

 

 

 

 

 

 

 

 

 

Variable Voltage

 

Range : 85-130 and 200-260 Volts

P, T

 

Auto Transformer

 

 

 

 

 

 

 

 

 

 

* P = Performance testing A = Calibration adjustments

T = Troubleshooting.

A-3

Page 17
Image 17
Agilent Technologies E3617A Performance Tests, Measurement Techniques, Type Required Characteristics USE Recommended Model