To Disconnect the Output Using an External Relay

Chapter 5 Service

To Disconnect the Output Using an External Relay

 

To Disconnect the Output Using an External Relay

 

When the output of the power supply is turned off, it is implemented by setting

 

the output to 0 volts and 0.02 amps. This gives a zero output voltage without

 

actually disconnecting the output. To disconnect the output, an external relay

 

must be connected between the output and the load. A TTL signal of either low

 

true or high true is provided to control an external relay. This signal can only

 

be controlled with the remote command OUTPut:RELay {OFFON}. The TTL

 

output is available on the RS-232 connector pin 1 and pin 9.

 

When the OUTPut:RELay state is ‘‘ON’’, the TTL output of pin 1 is high

 

(4.5 V) and pin 9 is low (0.5 V). The levels are reversed when the

 

OUTPut:RELay state is ‘‘OFF’’.

 

 

N o t e

TTL output of pin 1 or pin 9 of the RS-232 connector is available only after installing

 

two jumpers inside the power supply. See below for more information.

 

 

N o t e

Do not use the RS-232 interface if you have configured the power supply to output

 

relay control signals. Internal components on the RS-232 circuitry may be damaged.

 

Installation Procedure

 

 

The assembly drawings and schematics are located in chapter 8, ‘‘Schematics’’.

1Remove the front and rear bumpers and take off the cover (See the mechanical disassembly drawing on page 127).

2Install JP3 and JP4 located adjacent to the connector P5 (The JP3 and JP4 are outlined with a circle in the component locator diagram on page 128).

A bare wire may be used.

3Reassemble the power supply.