Chapter 3 Calibration Procedures

An Example program of Excel 97 for Calibration

ActiveCell.Value = "End Middle Current Calibration"

StartCalibration CurrentMax, False, shunt 'Set output to maximum current cal ActiveCell.Value = "End Maximum Current Calibration"

OVPandOCPCalibration False

Message = SendSCPI(power, "Syst:Err?")

If InStr(Message, "0") Then

ActiveCell.Value = "Current Calibration Complete"

Else

ActiveCell.Value = Message

ClosePort

Exit Sub

End If

EnableOVPandOCP True

SaveDate

ActiveCell.Value = "Calibration Complete"

Message = SendSCPI(power, "*RST")

ClosePort

End Sub

Private Function OpenPort() As Boolean 'Open communications on GPIB Dim Power_Address As String

Dim DMM_Address As String

Dim Error As Long

Power_Address = "5" 'Select power supply GPIB address between 0 to 30 DMM_Address = "22" 'Select DMM GPIB address between 0 to 30

Error = viOpenDefaultRM(id_power) 'Open the power supply VISA session If OpenPort = CheckError(Error, "Open ID Error to Power Supply") Then

Exit Function

End If

Error = viOpenDefaultRM(id_DMM) 'Open the DMM VISA session

If OpenPort = CheckError(Error, "Open ID Error to Digital Multi Meter") Then Exit Function

End If

Error = viOpen(id_power, "GPIB0::" & Power_Address & "::INSTR", 0, 1000, power) If OpenPort = CheckError(Error, "Unable Open to Power Supply") Then

Exit Function

End If

Error = viOpen(id_DMM, "GPIB0::" & DMM_Address & "::INSTR", 0, 1000, DMM) If OpenPort = CheckError(Error, "Unable Open to Digital Multi Meter") Then

Exit Function

End If

OpenPort = True

End Function