To Unsecure the Power Supply for Calibration

Error

Calibrate

I/O

Config

Secure

Chapter 3 Calibration Procedures

Calibration Security Code

To Unsecure the Power Supply for Calibration

The power supply can use a calibration security code to prevent unauthorized or accidental calibration. This procedure shows you how to unsecure the power supply for calibration from the front panel.

1Turn on the front-panel calibration mode.

SECURED

Turn on the calibration mode by holding down CalibrateError (Calibrate) key as you turn on the power supply and hold down the key until you hear a long beep.

If the power supply is secured, you will see the above message from the front panel for approximately one second. The ‘‘CAL MODE’’ message is then displayed on the front panel.

I/O

2 Move to the security code by pressing ConfigSecure (Secure) key.

000000 CODE

3Enter the security code using the knob and resolution selection keys.

003633 CODE (E3633A)

003634 CODE (E3634A)

The security code is set to ‘‘HP003633’’* or ‘‘HP003634’’** when the power supply is shipped from the factory. The security code is stored in non-volatile memory and does not change when the power has been off or after a remote interface reset.

To enter the security code from the front panel, enter only the last six digits.

To enter the security code from the remote interface, you may enter up to 12 characters.

Use the resolution selection keys to move left or right between digits. Use the knob to change the digits. Notice that the security code may be different if the security code has been changed from the default setting.

*For Agilent E3633A Model **For Agilent E3634A Model