Agilent Technologies E3634A, E3633A manual Front-Panel Voltage and Current Limit Settings

Models: E3634A E3633A

1 136
Download 136 pages 23.33 Kb
Page 5
Image 5
Front-Panel Voltage and Current Limit Settings

Front-Panel Voltage and Current Limit Settings

You can set the voltage and current limit values from the front panel using the following method.

Use the voltage/current adjust selection key, the resolution selection keys, and the control knob to change the voltage and current limit values.

1Select the desired range using the range selection keys after turning on the power supply.

2Press the DisplayLimit key to show the limit values on the display.

3Move the blinking digit to the appropriate position using the resolution

selection keys and change the blinking digit value to the desired voltage limit by turning the control knob. If the display limit times out, press the DisplayLimit key again.

4Set the knob to current control mode by pressing the CurrentVoltage key.

5Move the blinking digit to the appropriate position using the resolution selection keys and change the blinking digit value to the desired current limit by turning the control knob.

6Press the OutputOn/Off key to enable the output. After about 5 seconds, the display will go to output monitoring mode automatically to display the voltage and

current at the output or the display will go to output monitoring mode immediately by pressing the OutputOn/Off key again.

N o t e

All front panel keys and controls can be disabled with remote interface commands.

 

The Agilent E3633A and Agilent E3634A must be in "Local" mode for the front panel

 

keys and controls to function.

 

 

 

 

 

 

4

Page 5
Image 5
Agilent Technologies E3634A, E3633A manual Front-Panel Voltage and Current Limit Settings