Calibration Record for Agilent E3633A/E3634A

Chapter 3 Calibration Procedures

Calibration Record for Agilent E3633A/E3634A

Calibration Record for Agilent E3633A/E3634A

Step

Calibration Description

Measurement

Supply Being

Mode (DVM)

Adjusted

 

 

 

 

 

 

1

Unsecure the power supply (see page 62).

 

 

 

 

 

 

2

Turn on ‘‘CAL MODE’’ (hold down ‘‘Calibrate’’ and ‘‘Power’’ keys as

 

 

you turn on the power supply until you hear a long beeps).

 

 

 

 

 

 

 

 

 

3

Move down menu to ‘‘CAL SETUP 1’’ (press ‘‘Calibrate’’ key).

 

Voltage Calibration

 

 

 

 

 

Calibrate DAC and select the low point for voltage calibration; ‘‘START

 

 

 

BITCAL’’ appears for about 3 seconds and the display counts down

 

DAC and low voltage

4

numbers from 30 to 0. Then, ‘‘V LO 0.5000 V’’ appears on the display

V

point calibration

 

(press ‘‘Calibrate’’ key and wait about 30 seconds; then change the

 

 

 

display to match the DVM reading).

 

 

 

 

 

 

5

‘‘V MI 10.000 V’’* or ‘‘V MI 25.000 V’’** appears on the display (press

V

Middle voltage point

‘‘Calibrate’’ key; then change the display to match the DVM reading).

calibration

 

 

 

 

6

‘‘V HI 19.500 V’’* or ‘‘V HI 49.500 V’’** appears on the display (press

V

High voltage point

‘‘Calibrate’’ key; then change the display to match the DVM reading).

calibration

 

 

 

 

7

‘‘CAL SETUP 2’’ now appears on the display (press ‘‘Calibrate’’ key).

V

OVP calibration

 

 

 

 

8

‘‘CAL SETUP 3’’ now appears on the display (press ‘‘Calibrate’’ key;

 

Current calibration

then connect 0.01 9 resistor across the output terminals).

 

 

 

 

 

 

 

 

 

‘‘I LO 0.500 A’’* or ‘‘I LO 0.200 A’’** appears on the display (press

 

Low current point

9

‘‘Calibrate’’ key; then change the display to match the computed current

A

 

through the resistor).

 

calibration

 

 

 

 

 

 

 

 

‘‘I MI 10.00 A’’* or ‘‘I MI 3.500 A’’** appears on the display (press

 

Middle current point

10

‘‘Calibrate’’ key; then change the display to match the computed current

A

calibration

 

through the resistor).

 

 

 

 

 

 

 

‘‘I HI 19.5000 A’’* or ‘‘I HI 6.900 A’’** appears on the display (press

 

High current point

11

‘‘Calibrate’’ key; then change the display to match the computed current

A

calibration

 

through the resistor).

 

 

 

 

 

 

12

‘‘CAL SETUP 4’’ now appears on the display (press ‘‘Calibrate’’ key).

A

OCP calibration

 

 

 

 

13

Press ‘‘Calibrate’’ key, then press ‘‘Power’’ switch.

 

Exit CAL MODE

 

 

 

 

*For Agilent E3633A Model **For Agilent E3634A Model

72