Chapter 3 Calibration Procedures

An Example program of Excel 97 for Calibration

If DacErrorCorrection = False Then

ClosePort

Exit Sub

End If

shunt = 0.001 'Change the current shunt value

UserAnswer = MsgBox("Voltage & OVP calibration. Connect the output to the DMM", vbYesNo + vbQuestion, "E3633A Calibration")

If UserAnswer = vbNo Then

ClosePort

Exit Sub

End If

Range("B4").Select

ActiveCell.Value = "Begin Voltage Calibration"

StartCalibration VoltageMin, True, shunt 'Set output to minimum voltage cal ActiveCell.Value = "End Minimum Voltage Calibration"

StartCalibration VoltageMid, True, shunt 'Set output to middle voltage cal ActiveCell.Value = "End Middle Voltage Calibration"

StartCalibration VoltageMax, True, shunt 'Set output to maximum voltage cal ActiveCell.Value = "End Maximum Voltage Calibration"

Message = SendSCPI(power, "Syst:Err?")

If InStr(Message, "0") Then

ActiveCell.Value = "Voltage Calibration Complete"

Else

ActiveCell.Value = Message

ClosePort

Exit Sub

End If

OVPandOCPCalibration True

UserAnswer = MsgBox("Connect the Current shunt to the DMM input to measure a current.

Connect the output to the shunt.", vbYesNo + vbQuestion, "E3633A Calibration")

If UserAnswer = vbNo Then

ClosePort

Exit Sub

End If

ActiveCell.Value = "Begin Current Calibration"

StartCalibration CurrentMin,

False,

shunt

'Set

output to minimum current cal

ActiveCell.Value

= "End Minimum Current Calibration"

StartCalibration

CurrentMid,

False,

shunt

'Set

output to middle current cal

Continued on next page

76

Page 77
Image 77