•Adjust OCP trip level
Over Current

Over Current

Over Current

Over Current

Over Current

Chapter 2 Quick Start

To Program Overcurrent Protection

3Clear the overcurrent condition and exit this menu.

Now, when you press OverCurrent key again, the “DONE’’ message is displayed for a second and the OCP annunciator will not blink any more. The output will return to meter mode. The knob is selected for current control.

Notice that the power supply is operated in the constant current (CC) mode.

Adjust OCP trip level

1 Raise the OCP trip level.

Press OverCurrent key and turn the knob to raise the OCP trip level.

2Move to the OCP CLEAR mode.

OCP CLEAR

Press the Manual background key to move to the OCP CLEAR mode. The “OCP ON” message appears on the display. Turn the knob to the right until the above message appears on the display.

3Clear the overcurrent condition and exit this menu.

Now, when you press OverCurrent key again, the “DONE’’ message is displayed for a second and the OCP annunciator will not blink any more. The output will return to the meter mode.