Agilent Technologies E3633A, E3634A manual 1. Recommended Test Equipment

Models: E3634A E3633A

1 136
Download 136 pages 23.33 Kb
Page 48
Image 48
Chapter 3 Calibration Procedures

Chapter 3 Calibration Procedures

Recommended Test Equipment

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, use the accuracy requirements shown to select substitute calibration standards. If you use equipment other than that recommended in Table 3-1, you must recalculate the measurement uncertainties for the actual equipment used.

Table 3-1. Recommended Test Equipment

Instrument

Requirements

Recommended

Use

 

 

Model

 

 

 

 

 

GPIB controller

Full GPIB or RS-232 capabilities

Agilent 82341C

Programming and readback

 

 

interface card

accuracy

 

 

 

 

Oscilloscope

100 MHz with 20 MHz bandwidth

Agilent 54602B

Display transient response

 

 

 

and ripple & noise waveform

 

 

 

 

RMS Voltmeter

20 Hz to 20 MHz

 

Measure rms ripple & noise

 

 

 

 

Coaxial cable

 

Agilent 10502A

Measure rms ripple & noise

 

 

 

 

BNC female to

 

Agilent 1251-2277

Measure rms ripple & noise

banana plug adapter

 

 

 

 

 

 

 

Digital Voltmeter

Resolution: 0.1 mV

Agilent 34401A

Measure dc voltages

 

Accuracy: 0.01%

 

 

 

 

 

 

Electronic Load

Voltage Range: 60 Vdc

Agilent 60502B

Measure load and line

 

Current Range: 60 Adc

 

regulations and transient

 

 

response time.

 

Open and Short Switches

 

 

 

 

 

Transient On/Off

 

 

 

 

 

 

Resistive Loads

(0.4 9, 300 W/2.0 9, 300 W)*

 

Measure ripple and noise

(RL)

(3.5 9, 300 W/12.5 9, 300 W)**

 

 

Current monitoring

(0.01 9, 0.01%) [1]

ISOTEK Co.

Constant current test setup

Resistor (Shunt)

 

Model: A-H

 

 

 

 

 

 

 

 

3

[1]It is recommended to use a current monitoring resistor after calibration to find the accurate resistance.

*For Agilent E3633A Model **For Agilent E3634A Model

47