Agilent Technologies manual Performance Test Record for Agilent E3633A and E3634A

Models: E3634A E3633A

1 136
Download 136 pages 23.33 Kb
Page 61
Image 61
Performance Test Record for Agilent E3633A and E3634A

Chapter 3 Calibration Procedures

Performance Test Record for Agilent E3633A and E3634A

Performance Test Record for Agilent E3633A and E3634A

CV Performance Test Record

Test Description

Models

Actual Result

Specifications

 

 

Upper Limit

Lower Limit

 

 

 

 

 

 

 

 

CV Programming Accuracy @ 0

both

 

+0.0100 V

-0.0100 V

volts (DVM reading)

 

 

 

 

 

 

 

 

 

CV Readback Accuracy @ 0 volts

both

 

DVM + 0.0050 V

DVM - 0.0050 V

 

 

 

 

 

CV Programming Accuracy

(E3633A)

 

+20.0200 V

19.9800 V

@ Full Scale (DVM reading)

 

 

 

 

(E3634A)

 

+50.0350 V

49.9650 V

 

 

 

 

 

 

 

CV Readback Accuracy @ Full Scale

(E3633A)

 

DVM + 0.0150 V

DVM - 0.0150 V

 

 

 

 

 

 

(E3634A)

 

DVM + 0.0300 V

DVM - 0.0300 V

 

 

 

 

 

CV Load Effect (Load Regulation)

 

 

Maximum change: < (4 mV)* or (7 mV)**

 

 

 

 

CV Source Effect (Line Regulation)

 

 

Maximum change: < (4 mV)* or (7 mV)**

 

 

 

 

CV PARD (Normal mode)

 

 

< 3 mV p-p, (350 mV)* or (500 mV)** rms

 

 

 

 

Load Transient Response Time

both

 

< 50 msec

 

 

 

 

 

CC Performance Test Record

Test Description

Models

Actual Result

Specifications

 

 

Upper Limit

Lower Limit

 

 

 

 

 

 

 

 

 

 

CC Programming Accuracy

both

 

 

+0.0100 A

-0.0100 A

@ 0 amps (IO)

 

 

 

 

 

CC Readback Accuracy @ 0 amps

both

 

 

IO + 0.0050 A

IO - 0.0050 A

CC Programming Accuracy

(E3633A)

 

 

20.0500 A

19.9500 A

@ Full Scale (IO)

 

 

 

 

 

(E3634A)

 

 

7.0230 A

6.9770 A

 

 

 

 

 

 

CC Readback Accuracy @ Full Scale

(E3633A)

 

 

IO + 0.0350 A

IO - 0.0350 A

 

(E3634A)

 

 

IO + 0.0155 A

IO - 0.0155 A

CC Load Effect (Load Regulation)

 

 

 

Maximum change: < (2.25 mA)* or (0.95 mA)**

 

 

 

 

 

CC Source Effect (Line Regulation)

 

 

 

Maximum change: < (2.25 mA)* or (0.95 mA)**

 

 

 

 

 

CC PARD (Normal mode)

both

 

 

< 2 mA rms

 

 

 

 

 

CC PARD (Common mode)

both

 

 

< 1.5 mA rms

 

 

 

 

 

 

*For Agilent E3633A Model

**For Agilent E3634A Model

 